Set the voltage selectors on the rear panel of the separation and control
unit and development unit according to your local electricity supply;
110/120 V, or 220-230/240 V. To do this:
•
Check the voltage range of the mains electricity supply.
•
Set the voltage selector to the appropriate setting according to
the table below.
Voltage range
Voltage selector setting
For 120 V model instruments:
90-110
100
108-132
120
For 220 V model instruments:
198-242
220-230
216-264
240
Fuses
Each unit has two fuses. Check that the fuses are correctly installed
and intact.
Connecting the units
Connect the separation and control unit to the development unit with
the communication cable (code no. 19-6005-02).
Fig. 15.
PhastSystem controls rear
4. Installation