Symptoms
Solution(s)
Check the concentration of acetic
acid. It should be 5% in water.
Increase the time for this step.
Avoid touching the gel surface with
anything. Use the tab of gel backing
as a handle and use forceps to
handle the gel.
Check the concentration; use 0.5%
(or less) silver nitrate in solution.
1. Use fresh developer.
2. Develop gels for 1 or 2 minutes
longer.
3. Check that the gel holder rotates
during processing. If it dose not,
call service.
Probably Cause
Acetic acid wash ineffective in stop-
ping the developer.
Touching the gel with fingers or
metal objects.
Silver nitrate concentration is too
high.
1. Developer is too old.
2. The gels were not developed long
enough.
3. The gel did not rotate in the solu-
tions.
The gel
turns yellow
or brown
upon drying
Dark blotches
around gel edges
Very dark gel (black
or mirror effect)
Weakly stained
bands
Trouble Shooting Guide