4
119>CANNOT INSERT AT A RUN/RUNNING STEP
When trying to insert a step before, or at the
step in progress when you paused the run.
Steps can be inserted into a running method
only if they follow the paused running step.
120>ILLOGICAL TEMP. COMPENSATION CURVE
When starting a development run with a
method containing a temperature compensa-
tion (Ct) factor 0.0, or a Ct curve that de-
creases with increasing temperature. The tem-
perature compensation curve instruction i a
programming option: If you do not want tem-
perature compensation, make sure all the Ct
factors are set 1.0 for 5°C, 30°C, 40°C, and
50°C. See page 53 of the manual for more in-
formation.
121>MAX. STANDBY TEMP. IS 70°C
When trying to enter a standby temperature
greater than 70°C.
122>MAX. DEV TEMP. IS 50°
When trying to enter a development tempera-
ture greater than 50°C.
123>SEPARATION LID IS NOT CLOSED.
When trying ti start or continue a separation
run with the separation lid open.
During operation
If a system problem occurs during operation, the
running method will be paused automatically, an
alarm will sound, and one of the following
messages will appear on the display.
201>FAILED TO POSITION V3 TO EMPTY
202>FAILED TO POSITION V3 TO FILL
203>FAILED TO POSITION V3 TO NEUTRAL
204>FAILED TO POSITION V10 TO EMPTY
205>FAILED TO POSITION V10 TO FILL
206>FAILED TO POSITION V10 TO NEUTRAL
207>LEVEL SENSOR IN DEV CHAMBER NOT OK
If codes 201 to 207 appear on the display, call for
service. These messages appear when something
is wrong with the valves (V3 or V10) or the level
sensor in the development unit.
208>FAILED TO FILL CHAMBER – SEE MANUAL –
This message will appear if the chamber
cannot fill properly. This message can appear
even if the chamber is full. In this case, the
level sensor can not sense liquid in the
chamber. Try the following suggestions to
correct the problem:
•
Check if the chamber is full or empty.
Empty chamber (or semi-full chamber):
• Check bottles – see if they have enough liquid
in them and that the correct bottle is con-
nected to the port.
• Check tubing – see if there are any kinks or
punctures that prevent liquid flow to the de-
velopment chamber. Check tat they are sub-
merged in the liquid.
• Check the development lid – see if it is closed
tight, or if the gasket is loose or damaged. Re-
place the gasket if it is damaged. See page 78 of
the manual.
• If you suspect an obstruction in the tube be-
tween the 10-port valve and the development
chamber follow the instructions given for help
message 209 below, for unclogging ports.
Note: These instructions are for unclogging
the put-port, but the same instructions are
valid for unclogging the in-port.
• Dismantle and clean the 10-port valve accord-
ing to the instructions given on page 78 of the
manual. If this valve leaks, change the gasket.
To continue the run press ”DEV pause/con-
tinue”. The method will continue by ”FILLING”
the chamber.
Help Message Reference
Содержание PhastSystem
Страница 1: ...Phast System user manual automated electrophoresis um 80 1320 15 Edition AI ...
Страница 2: ......
Страница 8: ...8 ...
Страница 30: ...30 ...
Страница 34: ...34 ...
Страница 64: ...64 ...
Страница 96: ...12 ...
Страница 104: ......
Страница 105: ......
Страница 106: ...PRINTED IN SWEDEN BY TK I UPPSALA AB 2003 ...