49
Fig. 27.
Inserting the applicators.
Starting the run
1.
Press ’SEP start/stop” and enter the number of gels for this run:
NUMBER OF GELS
0
<do>
Methods are programmed for 1 gel. If you enter 2 gels here, the
current and power will be adjusted so that both gels run under
the same conditions according to the programmed method. See
page 14 for details.
2.
Enter the number of the method you plan to run:
START SEP Method 0.0
<do>
The method will start at the first step unless you renter a
different step number after the period.
Once you enter the method number, the method name will
appear in parentheses to the right of the method number (if you
gave your method a name), for example:
START SEP METHOD 4.0
(SDS-10000)
<do>
3.
Press ”do” to confirm.
Monitoring the run
If the separation bedtemperature (T) is warmer or cooler than the
programmed temperature (T
SET
) in the first step of the running
method, the display will show, for example:
SEP 4.1 COOLING BED T = 20°C T
SET
=18°C
SEP 4.1 HEATING BED T = 17°C T
SET
= 18°C
When T equals TSET the method will start and the running parameters
will appear on the display, for example:
SEP 4.1 400V 04.0mA 1.6W 18°C 0003A Vh
First the accumulated volthours (A Vh) are shown, that is, the number
of volthours that have elapsed since the beginning of the run. By
pressing ”SEP real condition” the number of elapsed volthours (Vh)
for the step which is running is displayed. Press ”SEP real condition” to
display A Vh again.
5. Operation
Содержание PhastSystem
Страница 1: ...Phast System user manual automated electrophoresis um 80 1320 15 Edition AI ...
Страница 2: ......
Страница 8: ...8 ...
Страница 30: ...30 ...
Страница 34: ...34 ...
Страница 64: ...64 ...
Страница 96: ...12 ...
Страница 104: ......
Страница 105: ......
Страница 106: ...PRINTED IN SWEDEN BY TK I UPPSALA AB 2003 ...