CARESCAPE Monitor B850
156
2106778-001
12.2 Unit disassembly
12.2.1 Opening the enclosure cover
When the CPU unit is opened, the electrical safety tests must always be performed after the
monitor is reassembled. For more information, refer to the
requirements in the
chapter.
1.
Remove 4 screws (T10) on each side of the top cover. Torque [1.1 Nm].
2.
Remove the top cover.
12.2.2 Replacing the CPU timekeeper battery
1.
Open the unit as described in
.
2.
Remove the lithium ion battery from the bracket on the processor board.
3.
Replace the battery.
4.
Reassemble in reverse order.
To avoid network device time synchronization issues, remember to readjust the time and date
before you connect the monitor to network. For instructions on configuring the time, see
Содержание CARESCAPE Monitor B850
Страница 12: ...CARESCAPE Monitor B850 4 2106778 001 For your notes ...
Страница 16: ...CARESCAPE Monitor B850 8 2106778 001 For your notes ...
Страница 56: ...CARESCAPE Monitor B850 48 2106778 001 For your notes ...
Страница 76: ...CARESCAPE Monitor B850 68 2106778 001 For your notes ...
Страница 112: ...CARESCAPE Monitor B850 104 2106778 001 For your notes ...
Страница 128: ...CARESCAPE Monitor B850 120 2106778 001 For your notes ...
Страница 156: ...CARESCAPE Monitor B850 148 2106778 001 ...
Страница 157: ...Troubleshooting 149 2106778 001 Problem No waveforms or parameters are displayed at the CARESCAPE Cental Station ...
Страница 160: ...CARESCAPE Monitor B850 152 2106778 001 For your notes ...
Страница 177: ...Service parts 169 2106778 001 13 3 Exploded view ...
Страница 178: ...CARESCAPE Monitor B850 170 2106778 001 For your notes ...
Страница 182: ...CARESCAPE Monitor B850 A 4 4 2106778 001 Notes Used service parts Signature Date ...
Страница 186: ...CARESCAPE Monitor B850 B 4 4 2106778 001 For your notes ...
Страница 190: ...CARESCAPE Monitor B850 C 4 4 2106778 001 For your notes ...
Страница 191: ......