CARESCAPE Monitor B850
78
2106778-001
2.
Select
Save
.
If you selected
No Parser
, the barcode setting configuration is complete.
For a
Length
or
Character Delimited Parser
, follow the applicable instructions.
7.8.3 Configuring length delimited parser information
.
or
7.8.4 Configure character delimited parser information
.
7.8.3 Configuring length delimited parser information
Points to note
If you configure
Age
, you must either select the
Age Unit
item or one of the age units (e.g.,
Years, Months, Weeks, Days) below
Fixed Option
.
If you configure
Height
, you must either select the
Height Unit
item or one of the height
units (e.g., Feet, Inches, Meters, Centimeters, Millimeters) below
Fixed Option
.
If you configure
Weight
, you must either select the
Weight Unit
item or one of the weight
units (e.g., Kilograms, Grams, Micrograms, Pounds, Ounces) below
Fixed Option
.
For an example of admit/discharge configuration for a length delimited parser, see
Example of length delimited parser information on page 79.
1.
In the
Admit/Discharge Configuration
window, enter the location and length information
for each data item.
If an item is not included in the barcode, type 0 in the item’s
Position
and
Length
fields, or
leave the
Position
and
Length
fields blank.
a.
In the
Position
column, type the beginning position of the field in the data string
(from 1 to 300).
b.
In the
Length
column, type the number of characters (from 1 to 99) that the field
contains.
2.
For
Gender Format
, select
Fixed
or
Configured
.
If you select
Configured
:
a.
Type the character that identifies
Male
.
b.
Type the character that identifies
Female
.
3.
Below
Fixed Option
, select the applicable value:
4.
Scroll to the bottom of the window, and select
Save
.
All changes take effect immediately.
Item
Item selection in the top part of
the screen
Fixed Option selection
Height Unit
Both
Height
and
Height Unit
Non-Fixed
.
Height
only
Select value from drop-down list.
Weight Unit
Both
Weight
and
Weight Unit
Non-Fixed
.
Weight
only
Select value from drop-down list.
Age Unit
Both
Age
and
Age Unit
Non-Fixed
.
Age
only
Select value from drop-down list.
Содержание CARESCAPE Monitor B850
Страница 12: ...CARESCAPE Monitor B850 4 2106778 001 For your notes ...
Страница 16: ...CARESCAPE Monitor B850 8 2106778 001 For your notes ...
Страница 56: ...CARESCAPE Monitor B850 48 2106778 001 For your notes ...
Страница 76: ...CARESCAPE Monitor B850 68 2106778 001 For your notes ...
Страница 112: ...CARESCAPE Monitor B850 104 2106778 001 For your notes ...
Страница 128: ...CARESCAPE Monitor B850 120 2106778 001 For your notes ...
Страница 156: ...CARESCAPE Monitor B850 148 2106778 001 ...
Страница 157: ...Troubleshooting 149 2106778 001 Problem No waveforms or parameters are displayed at the CARESCAPE Cental Station ...
Страница 160: ...CARESCAPE Monitor B850 152 2106778 001 For your notes ...
Страница 177: ...Service parts 169 2106778 001 13 3 Exploded view ...
Страница 178: ...CARESCAPE Monitor B850 170 2106778 001 For your notes ...
Страница 182: ...CARESCAPE Monitor B850 A 4 4 2106778 001 Notes Used service parts Signature Date ...
Страница 186: ...CARESCAPE Monitor B850 B 4 4 2106778 001 For your notes ...
Страница 190: ...CARESCAPE Monitor B850 C 4 4 2106778 001 For your notes ...
Страница 191: ......