8.2.8 Patient (source) leakage current tests
This procedure measures the leakage current from an applied part connector of the device to
ground.
Perform the test in Normal Condition (NC) and in two different Single Fault Conditions (SFC):
1) earth open and 2) one of the supply conductors open at a time.
Perform test with normal and reverse polarity.
NOTE: Refer to the instructions delivered with the safety analyzer to perform this test.
NOTE: *The measuring device (MD) represents the network and voltage measuring instrument
and its frequency characters according to IEC 60601-1.
E-COP, E-COPSv
P4/P8
E-PiCCO
P8
E-NSATX, E-NSAT
SpO2
E-Masimo
SpO2
E-NMT
NMT
E-BIS
1. Connect the BISx Digital Signal Processing Unit with the Patient
Interface Cable (PIC+) to the E-BIS module.
2. Connect the specified test body to the PIC+ cable.
NOTE: The patient isolation is in the BISx Digital Signal Processing
Unit, not in the E-BIS module.
E-Entropy
1. Connect an Entropy sensor cable to the module.
2. Connect the specified test body to the Entropy sensor cable.
E-EEG 1. Disconnect
the N-EEG headbox from the E-EEG module.
2. Connect the test body directly to the E-EEG module.
PDM
ECG &
SpO2
SpO2, Masimo SpO2
SpO2
Module
Patient connector
Содержание CARESCAPE Monitor B850
Страница 12: ...CARESCAPE Monitor B850 4 2106778 001 For your notes ...
Страница 16: ...CARESCAPE Monitor B850 8 2106778 001 For your notes ...
Страница 56: ...CARESCAPE Monitor B850 48 2106778 001 For your notes ...
Страница 76: ...CARESCAPE Monitor B850 68 2106778 001 For your notes ...
Страница 112: ...CARESCAPE Monitor B850 104 2106778 001 For your notes ...
Страница 128: ...CARESCAPE Monitor B850 120 2106778 001 For your notes ...
Страница 156: ...CARESCAPE Monitor B850 148 2106778 001 ...
Страница 157: ...Troubleshooting 149 2106778 001 Problem No waveforms or parameters are displayed at the CARESCAPE Cental Station ...
Страница 160: ...CARESCAPE Monitor B850 152 2106778 001 For your notes ...
Страница 177: ...Service parts 169 2106778 001 13 3 Exploded view ...
Страница 178: ...CARESCAPE Monitor B850 170 2106778 001 For your notes ...
Страница 182: ...CARESCAPE Monitor B850 A 4 4 2106778 001 Notes Used service parts Signature Date ...
Страница 186: ...CARESCAPE Monitor B850 B 4 4 2106778 001 For your notes ...
Страница 190: ...CARESCAPE Monitor B850 C 4 4 2106778 001 For your notes ...
Страница 191: ......