The quality of the display is not
good.
•
The cable is not connected
properly.
•
The video display is bad.
1. Check the connection between the
video display and the patient monitor.
If necessary, replace the video display
with a known good one.
2. If the problem is not corrected, replace
the processor board.
The power on/off switch on the
patient monitor does not
illuminate when it is toggled.
The power supply is bad.
Replace the power supply.
Continuous beeping alarm and
alarm light flashing yellow.
Mains supply is lost or the USB
cable is disconnected.
Restore the mains supply or reconnect the
USB cable.
Problem
Possible cause
Recommended actions
Touchscreen operation
inaccurate.
Touchscreen not calibrated.
Calibrate touchscreen, see section
.
Touchscreen inoperative.
Touchscreen USB cable loose
Connect the USB touchscreen cable to the
CPU unit or to the display.
NOTE: Secure the connected USB cable to
the CPU unit. Use the existing retaining
clips attached to the CPU unit.
Touchscreen USB cable
connected to the wrong USB
connector in the monitor.
Connect the touchscreen USB cable to the
correct connector.
Faulty touchscreen sensor.
Replace the display.
Problem
Possible cause
Recommended actions
Содержание CARESCAPE Monitor B850
Страница 12: ...CARESCAPE Monitor B850 4 2106778 001 For your notes ...
Страница 16: ...CARESCAPE Monitor B850 8 2106778 001 For your notes ...
Страница 56: ...CARESCAPE Monitor B850 48 2106778 001 For your notes ...
Страница 76: ...CARESCAPE Monitor B850 68 2106778 001 For your notes ...
Страница 112: ...CARESCAPE Monitor B850 104 2106778 001 For your notes ...
Страница 128: ...CARESCAPE Monitor B850 120 2106778 001 For your notes ...
Страница 156: ...CARESCAPE Monitor B850 148 2106778 001 ...
Страница 157: ...Troubleshooting 149 2106778 001 Problem No waveforms or parameters are displayed at the CARESCAPE Cental Station ...
Страница 160: ...CARESCAPE Monitor B850 152 2106778 001 For your notes ...
Страница 177: ...Service parts 169 2106778 001 13 3 Exploded view ...
Страница 178: ...CARESCAPE Monitor B850 170 2106778 001 For your notes ...
Страница 182: ...CARESCAPE Monitor B850 A 4 4 2106778 001 Notes Used service parts Signature Date ...
Страница 186: ...CARESCAPE Monitor B850 B 4 4 2106778 001 For your notes ...
Страница 190: ...CARESCAPE Monitor B850 C 4 4 2106778 001 For your notes ...
Страница 191: ......