CARESCAPE Monitor B850
92
2106778-001
6.
Select
Activate
to start the settings activation process.
•
If immediate settings activation was selected:
a.
Wait until the settings activation is completed and the patient monitor has
performed an automatic restart.
b.
Verify that the settings activation was successful and the patient monitor is running
the activated settings.
•
If after next case end / discharge settings activation was selected:
a.
The patient monitor will show a message
Settings activation on next
discharge/after next case end
on the display until the next case end / patient
discharge. The patient monitoring can be continued normally until then.
b.
Settings activation will start automatically after the next case end / patient
discharge, unless this process is canceled by the user.
c.
Wait until the settings activation is completed and the patient monitor has
performed an automatic restart.
d.
Verify that the settings activation was successful and the patient monitor is running
the activated settings.
7.17.4 Canceling pending settings activation
You can cancel pending settings activation anytime while the message
Setting activation
after next case end / Setting activation after next discharge
is shown on the display.
1.
Log in to Webmin.
2.
Select
Configuration
>
Settings
.
3.
In the
Sub-Modules for Settings
menu, select
Activate
.
4.
Select
Cancel
to cancel settings activation.
Select
To
NOTE
Immediately
Activate the new settings
Immediately.
The patient monitor must be in a
case reset / patient discharged state.
After next case end /
discharge
Take the new settings in use
with a delay, after the next
case end / patient discharge.
This option allows you to initiate the
setting activation process with a
delay during an active patient case.
The new settings will activate only
after the next case end / patient
discharge.
Содержание CARESCAPE Monitor B850
Страница 12: ...CARESCAPE Monitor B850 4 2106778 001 For your notes ...
Страница 16: ...CARESCAPE Monitor B850 8 2106778 001 For your notes ...
Страница 56: ...CARESCAPE Monitor B850 48 2106778 001 For your notes ...
Страница 76: ...CARESCAPE Monitor B850 68 2106778 001 For your notes ...
Страница 112: ...CARESCAPE Monitor B850 104 2106778 001 For your notes ...
Страница 128: ...CARESCAPE Monitor B850 120 2106778 001 For your notes ...
Страница 156: ...CARESCAPE Monitor B850 148 2106778 001 ...
Страница 157: ...Troubleshooting 149 2106778 001 Problem No waveforms or parameters are displayed at the CARESCAPE Cental Station ...
Страница 160: ...CARESCAPE Monitor B850 152 2106778 001 For your notes ...
Страница 177: ...Service parts 169 2106778 001 13 3 Exploded view ...
Страница 178: ...CARESCAPE Monitor B850 170 2106778 001 For your notes ...
Страница 182: ...CARESCAPE Monitor B850 A 4 4 2106778 001 Notes Used service parts Signature Date ...
Страница 186: ...CARESCAPE Monitor B850 B 4 4 2106778 001 For your notes ...
Страница 190: ...CARESCAPE Monitor B850 C 4 4 2106778 001 For your notes ...
Страница 191: ......