CARESCAPE Monitor B850
A - 2 (4)
2106778-001
Yes:
Yes: you do not have to perform the electrical safety tests. To continue the
installation check, mark the electrical safety test completion N.A. and
proceed to
. If available, attach the original electrical safety
test results to this check form.
No:
No: continue to perform the electrical safety tests.
Power outlet is correctly wired.
Power cord and plug are undamaged and all conductors are properly connected.
Observed result
Acceptance
criteria
PASS
N.A.
FAIL
a.) Ground continuity test
without power cord
0.1 ohms
with power cord
0.2 ohms
b.) Impedance of protective earth
connection
without power cord
0.1 ohms
with power cord
0.2 ohms
Normal Condition (NC)
500 µA
EN /IEC
300 µA
UL
Single Fault Condition (SFC)
1 mA
Normal Condition (NC)
100 µA
Single Fault Condition (SFC)
500 µA
EN /IEC
300 µA
UL
Module type
S/N
Observed result
Acceptance
criteria
PASS
N.A.
FAIL
ECG
SpO2/
InvP
Patient (source) leakage current tests, using a test body
Normal Condition (NC)
Polarity: NORMAL
10 µA
Single Fault Condition (SFC)
Ground closed (normal)
50 µA
Ground open (normal)
50 µA
Ground closed (reverse)
50 µA
Ground open (reverse)
50 µA
Normal Condition (NC)
Polarity: Reverse
10 µA
Patient (sink) leakage current tests
Polarity: NORMAL
50
µA
Polarity: Reverse
50
µA
Module type
S/N
Observed result
Acceptance
criteria
PASS
N.A.
FAIL
Patient (source) leakage current tests, using a test body
Normal Condition (NC)
Polarity: NORMAL
10 µA
Содержание CARESCAPE Monitor B850
Страница 12: ...CARESCAPE Monitor B850 4 2106778 001 For your notes ...
Страница 16: ...CARESCAPE Monitor B850 8 2106778 001 For your notes ...
Страница 56: ...CARESCAPE Monitor B850 48 2106778 001 For your notes ...
Страница 76: ...CARESCAPE Monitor B850 68 2106778 001 For your notes ...
Страница 112: ...CARESCAPE Monitor B850 104 2106778 001 For your notes ...
Страница 128: ...CARESCAPE Monitor B850 120 2106778 001 For your notes ...
Страница 156: ...CARESCAPE Monitor B850 148 2106778 001 ...
Страница 157: ...Troubleshooting 149 2106778 001 Problem No waveforms or parameters are displayed at the CARESCAPE Cental Station ...
Страница 160: ...CARESCAPE Monitor B850 152 2106778 001 For your notes ...
Страница 177: ...Service parts 169 2106778 001 13 3 Exploded view ...
Страница 178: ...CARESCAPE Monitor B850 170 2106778 001 For your notes ...
Страница 182: ...CARESCAPE Monitor B850 A 4 4 2106778 001 Notes Used service parts Signature Date ...
Страница 186: ...CARESCAPE Monitor B850 B 4 4 2106778 001 For your notes ...
Страница 190: ...CARESCAPE Monitor B850 C 4 4 2106778 001 For your notes ...
Страница 191: ......