Configuration
75
2106778-001
4.
Select
Save
.
All changes take effect immediately.
Citrix Session Timeout (in Minutes)
Username
Password
Confirm Password
NOTE: The hospital’s IT Administrator or biomedical department can supply these values.
5.
Select
Save
.
All changes take effect immediately.
7.6.1 Troubleshooting
The information entered must confirm to the following standards:
Citrix Server Address: Any valid IPv4 address.
Initial Program: Name of Citrix-published application to launch.
Citrix User Name: Any valid Windows username. If blank, the user will be forced to enter it
manually.
Citrix Password: Any valid Windows password.
Citrix Timeout: Any integer greater than or equal to 0 and less than 9999.
Citrix Startup Time: Any integer greater than or equal to 0 and less than 9999.
7.6.2 Glossary
Citrix Server Address: IP address or DNS name of Citrix Server.
Initial Program:
Name of Citrix-published application to launch.
Citrix User Name:
ID used to log into Citrix session.
Citrix Password:
Password used to log into Citrix session.
Username
Any valid Windows username. You can leave
this field empty to allow anonymous login.
The maximum length is 128 characters.
Password
Any valid Windows password. You can leave
this field empty to allow anonymous login.
The maximum length is 128 characters.
Confirm Password
The password and confirmed password must
be the same.
Encryption Level
Select one of the following encryption levels:
- "Basic"
- "RC5 (128 bit - Login Only)"
- "RC5 (40 bit)"
- "RC5 (56 bit)"
- "RC5 (128 bit)"
Field name
Description and valid choices
Содержание CARESCAPE Monitor B850
Страница 12: ...CARESCAPE Monitor B850 4 2106778 001 For your notes ...
Страница 16: ...CARESCAPE Monitor B850 8 2106778 001 For your notes ...
Страница 56: ...CARESCAPE Monitor B850 48 2106778 001 For your notes ...
Страница 76: ...CARESCAPE Monitor B850 68 2106778 001 For your notes ...
Страница 112: ...CARESCAPE Monitor B850 104 2106778 001 For your notes ...
Страница 128: ...CARESCAPE Monitor B850 120 2106778 001 For your notes ...
Страница 156: ...CARESCAPE Monitor B850 148 2106778 001 ...
Страница 157: ...Troubleshooting 149 2106778 001 Problem No waveforms or parameters are displayed at the CARESCAPE Cental Station ...
Страница 160: ...CARESCAPE Monitor B850 152 2106778 001 For your notes ...
Страница 177: ...Service parts 169 2106778 001 13 3 Exploded view ...
Страница 178: ...CARESCAPE Monitor B850 170 2106778 001 For your notes ...
Страница 182: ...CARESCAPE Monitor B850 A 4 4 2106778 001 Notes Used service parts Signature Date ...
Страница 186: ...CARESCAPE Monitor B850 B 4 4 2106778 001 For your notes ...
Страница 190: ...CARESCAPE Monitor B850 C 4 4 2106778 001 For your notes ...
Страница 191: ......