Configuration
99
2106778-001
4.
In
Erase Inactive Software After Activation
, select:
5.
Select
Activate
to start the host software activation.
6.
Confirm activation by selecting
Yes
in the Host Software window that opens.
The patient monitor shows a screen saver that informs about the ongoing software
activation:
Software activation in progress. Do not disconnect any measurement modules or
other peripheral devices, or shut down the monitor until the software activation is
complete. Activation may take up to 10 minutes. The device will automatically restart
once the software activation is complete.
NOTE: When the software is activated first time after a uDOM replacement, the screen
saver does not appear.
7.
Wait until the software activation completes and the patient monitor restarts
automatically.
8.
Verify that the software activation is successful and the patient monitor runs the
activated software.
Activating host software after next case end / discharge
If you want to activate the new host software with a time delay, that is, after the next case end
/ patient discharge, follow this instruction. You can initiate the software activation process with
a delay during an active patient case.
No
- to keep the currently active software version as inactive software after
the new software version is activated successfully. This option lets you
restore the patient monitor to the previous software version later.
Yes
- to erase the currently active software version permanently after the
new software version is activated successfully.
Содержание CARESCAPE Monitor B850
Страница 12: ...CARESCAPE Monitor B850 4 2106778 001 For your notes ...
Страница 16: ...CARESCAPE Monitor B850 8 2106778 001 For your notes ...
Страница 56: ...CARESCAPE Monitor B850 48 2106778 001 For your notes ...
Страница 76: ...CARESCAPE Monitor B850 68 2106778 001 For your notes ...
Страница 112: ...CARESCAPE Monitor B850 104 2106778 001 For your notes ...
Страница 128: ...CARESCAPE Monitor B850 120 2106778 001 For your notes ...
Страница 156: ...CARESCAPE Monitor B850 148 2106778 001 ...
Страница 157: ...Troubleshooting 149 2106778 001 Problem No waveforms or parameters are displayed at the CARESCAPE Cental Station ...
Страница 160: ...CARESCAPE Monitor B850 152 2106778 001 For your notes ...
Страница 177: ...Service parts 169 2106778 001 13 3 Exploded view ...
Страница 178: ...CARESCAPE Monitor B850 170 2106778 001 For your notes ...
Страница 182: ...CARESCAPE Monitor B850 A 4 4 2106778 001 Notes Used service parts Signature Date ...
Страница 186: ...CARESCAPE Monitor B850 B 4 4 2106778 001 For your notes ...
Страница 190: ...CARESCAPE Monitor B850 C 4 4 2106778 001 For your notes ...
Страница 191: ......