Installation checkout
115
2106778-001
7.
Repeat this test for all the connected parameter modules and patient connectors
specified in
Acceptance criteria:
•
All readings shall be less than or equal to 50
A.
8.2.10 Test completion
1.
Disconnect the safety analyzer from the power outlet.
2.
Disconnect the test equipment from the patient monitor.
3.
Disconnect the patient monitor’s power cord from the leakage tester.
4.
Mark this task as complete on the Appendix
.
8.3 Functional check
The purpose of this functional check is to ensure that the system is properly installed and
configured. Cover all peripheral devices that are connected to the patient monitor by
performing the applicable tests below. Skip the tests that are not applicable for the installed
patient monitor.
8.3.1 Start-up
1.
Connect the power cord to the wall outlet.
2.
Turn on the patient monitor.
Verify that the patient monitor starts up normally:
The speaker gives an audible beep.
The normal monitoring screen appears and there are no error messages on the
screen.
8.3.2 Display
Picture quality
Perform this test for all the connected displays .
1.
Verify that all text is readable and all images are clear.
2.
Verify that the brightness is good. Adjust if necessary.
Touchscreen control
Perform this test for all connected touchscreen displays.
1.
Verify the operation and the calibration of a touchscreen by touching a corner of an
active parameter window. Verify that the selected menu is opened.
2.
Select
to close any open menu and return to the main display.
Alarm light
Perform this test for all the connected displays.
1.
Create an active alarm and check that the alarm light illuminates.
8.3.3 Device Information
1.
Log in to Webmin.
Содержание CARESCAPE Monitor B850
Страница 12: ...CARESCAPE Monitor B850 4 2106778 001 For your notes ...
Страница 16: ...CARESCAPE Monitor B850 8 2106778 001 For your notes ...
Страница 56: ...CARESCAPE Monitor B850 48 2106778 001 For your notes ...
Страница 76: ...CARESCAPE Monitor B850 68 2106778 001 For your notes ...
Страница 112: ...CARESCAPE Monitor B850 104 2106778 001 For your notes ...
Страница 128: ...CARESCAPE Monitor B850 120 2106778 001 For your notes ...
Страница 156: ...CARESCAPE Monitor B850 148 2106778 001 ...
Страница 157: ...Troubleshooting 149 2106778 001 Problem No waveforms or parameters are displayed at the CARESCAPE Cental Station ...
Страница 160: ...CARESCAPE Monitor B850 152 2106778 001 For your notes ...
Страница 177: ...Service parts 169 2106778 001 13 3 Exploded view ...
Страница 178: ...CARESCAPE Monitor B850 170 2106778 001 For your notes ...
Страница 182: ...CARESCAPE Monitor B850 A 4 4 2106778 001 Notes Used service parts Signature Date ...
Страница 186: ...CARESCAPE Monitor B850 B 4 4 2106778 001 For your notes ...
Страница 190: ...CARESCAPE Monitor B850 C 4 4 2106778 001 For your notes ...
Страница 191: ......