Pre-installation requirements
49
2106778-001
5
Pre-installation requirements
This chapter specifies the pre-installation requirements for the patient monitor.
5.1 Unpacking
1.
Confirm that the packing box is undamaged. If the box is damaged, contact the shipper.
2.
Open the top of the box and carefully unpack all components.
3.
Confirm that all components are undamaged. If any of the components is damaged,
contact the shipper.
4.
Confirm that all components are included. If any of the components is missing, contact
your GE Healthcare distributor.
WARNING
EXCESSIVE LEAKAGE CURRENT - If the device has been transported or stored
outside operating temperature range allow it to stabilize back to operating
temperature range before removing it from the plastic bag.
CAUTION
PACKAGING DISPOSAL - Dispose of the packaging material, observing the
applicable waste control regulations.
5.2 Compatibility check
Verify the compatibility of all system components prior to the installation of the patient
monitor.
•
Refer to the patient monitor’s supplemental information manual for a list of compatible
displays and display cables.
•
Refer to the patient monitor’s supplemental information manual for a list of compatible
network and bedside devices.
•
Refer to the patient monitor’s supplemental information manual for a list of compatible
supplies and accessories.
•
Refer to the patient monitor’s supplemental information manual and Unity Network
Interface Device (ID) Operator's Manual to see compatible peripheral devices.
WARNING
BEFORE INSTALLATION - Compatibility is critical to safe and effective use of
this device. Please contact your local sales or service representative prior to
installation to verify equipment compatibility.
WARNING
INTERFACING OTHER EQUIPMENT - Connect only items that are specified as
part of the system and as compatible. For more information, see the
CARESCAPE Modular Monitors Supplemental Information Manual.
WARNING
Before connecting an interfacing module to the device, verify compatibility.
Verify the connectivity of device interfaces before using the equipment.
Verify the compatibility of software versions before using the equipment.
WARNING
Do not use identical measurement modules or modules that map a
measurement to the same channel or parameter window. If such modules
have been connected, remove the module that has been most recently
connected. You can also remove both modules and re-connect the new
module after five seconds.
Содержание CARESCAPE Monitor B850
Страница 12: ...CARESCAPE Monitor B850 4 2106778 001 For your notes ...
Страница 16: ...CARESCAPE Monitor B850 8 2106778 001 For your notes ...
Страница 56: ...CARESCAPE Monitor B850 48 2106778 001 For your notes ...
Страница 76: ...CARESCAPE Monitor B850 68 2106778 001 For your notes ...
Страница 112: ...CARESCAPE Monitor B850 104 2106778 001 For your notes ...
Страница 128: ...CARESCAPE Monitor B850 120 2106778 001 For your notes ...
Страница 156: ...CARESCAPE Monitor B850 148 2106778 001 ...
Страница 157: ...Troubleshooting 149 2106778 001 Problem No waveforms or parameters are displayed at the CARESCAPE Cental Station ...
Страница 160: ...CARESCAPE Monitor B850 152 2106778 001 For your notes ...
Страница 177: ...Service parts 169 2106778 001 13 3 Exploded view ...
Страница 178: ...CARESCAPE Monitor B850 170 2106778 001 For your notes ...
Страница 182: ...CARESCAPE Monitor B850 A 4 4 2106778 001 Notes Used service parts Signature Date ...
Страница 186: ...CARESCAPE Monitor B850 B 4 4 2106778 001 For your notes ...
Страница 190: ...CARESCAPE Monitor B850 C 4 4 2106778 001 For your notes ...
Страница 191: ......