CARESCAPE Monitor B850
134
2106778-001
11.3.2 Ping a TCP/IP network device
Use this Webmin feature to verify connectivity with a network device on the MC Network and IX
Network.
1.
Log in to Webmin.
2.
Select
Diagnostics > Ping
.
3.
In the
Address to Ping
field in the Ping Command window, type the IP address of a known
device on the network and select ping.
If you receive a reply, then you are able to connect to the device.
If you do not receive a reply, make sure that the patient monitor is connected to an active
network.
The patient monitor withstands a maximum packet loss of 5 packets per 1 million and
maximum latency of 250 ms without performance degradation.
11.3.3 Log files
The patient monitor collects information about different system events and errors to log files.
These log files help troubleshooting problems in the patient monitor and the connected
peripheral devices.
The following table describes the available log files and the type of information that they
collect.
battery (in mV)
Lithium Battery Voltage for the Real Time Clock
Ambient temperature (in C)
Ambient temperature of the CPU unit measured by the
sensor in the main processor board
CPU temperature (in C)
An internal temperature of the CPU unit measured by
the sensor in the main processor
945 temperature (in C)
An internal temperature of the CPU unit measured by
the sensor in the Intel 945 Graphic controller chipset (1st
and 2rd display port) temperature
VideoCard temperature (in C)
An internal temperature of the CPU unit measured by
the sensor in the additional 3rd video board (optional)
Measurement
Description
Содержание CARESCAPE Monitor B850
Страница 12: ...CARESCAPE Monitor B850 4 2106778 001 For your notes ...
Страница 16: ...CARESCAPE Monitor B850 8 2106778 001 For your notes ...
Страница 56: ...CARESCAPE Monitor B850 48 2106778 001 For your notes ...
Страница 76: ...CARESCAPE Monitor B850 68 2106778 001 For your notes ...
Страница 112: ...CARESCAPE Monitor B850 104 2106778 001 For your notes ...
Страница 128: ...CARESCAPE Monitor B850 120 2106778 001 For your notes ...
Страница 156: ...CARESCAPE Monitor B850 148 2106778 001 ...
Страница 157: ...Troubleshooting 149 2106778 001 Problem No waveforms or parameters are displayed at the CARESCAPE Cental Station ...
Страница 160: ...CARESCAPE Monitor B850 152 2106778 001 For your notes ...
Страница 177: ...Service parts 169 2106778 001 13 3 Exploded view ...
Страница 178: ...CARESCAPE Monitor B850 170 2106778 001 For your notes ...
Страница 182: ...CARESCAPE Monitor B850 A 4 4 2106778 001 Notes Used service parts Signature Date ...
Страница 186: ...CARESCAPE Monitor B850 B 4 4 2106778 001 For your notes ...
Страница 190: ...CARESCAPE Monitor B850 C 4 4 2106778 001 For your notes ...
Страница 191: ......