CARESCAPE Monitor B850
100
2106778-001
1.
In the
Host Software
window, verify that the software that you activate is listed as
Inactive
in
Current Software State
.
2.
In
Software License
, if applicable, enter the
Activation Code
for the new software
version.
3.
In
Software Activation
, select
After next case end / discharge.
4.
In
Erase Inactive Software After Activation
, select:
5.
Select
Activate
to start the software activation.
6.
Confirm activation by selecting
Yes
in the
Host Software
window that opens.
The patient monitor shows a message
Software activation after next case end / after
next discharge
on the display until the next case end or patient discharge. The patient
monitoring can be continued normally until then.
7.
Software activation will start automatically after the next case end / patient discharge,
unless this process is cancelled before it starts.
The patient monitor informs clinical users about pending software activation by showing the
message
Software activation after next case end / after next discharge
on the display.
No
- to keep the currently active software version as inactive software after
the new software version is activated successfully. This option lets you
restore the patient monitor to the previous software version later.
Yes
- to erase the currently active software version permanently after the
new software version is successfully activated.
Содержание CARESCAPE Monitor B850
Страница 12: ...CARESCAPE Monitor B850 4 2106778 001 For your notes ...
Страница 16: ...CARESCAPE Monitor B850 8 2106778 001 For your notes ...
Страница 56: ...CARESCAPE Monitor B850 48 2106778 001 For your notes ...
Страница 76: ...CARESCAPE Monitor B850 68 2106778 001 For your notes ...
Страница 112: ...CARESCAPE Monitor B850 104 2106778 001 For your notes ...
Страница 128: ...CARESCAPE Monitor B850 120 2106778 001 For your notes ...
Страница 156: ...CARESCAPE Monitor B850 148 2106778 001 ...
Страница 157: ...Troubleshooting 149 2106778 001 Problem No waveforms or parameters are displayed at the CARESCAPE Cental Station ...
Страница 160: ...CARESCAPE Monitor B850 152 2106778 001 For your notes ...
Страница 177: ...Service parts 169 2106778 001 13 3 Exploded view ...
Страница 178: ...CARESCAPE Monitor B850 170 2106778 001 For your notes ...
Страница 182: ...CARESCAPE Monitor B850 A 4 4 2106778 001 Notes Used service parts Signature Date ...
Страница 186: ...CARESCAPE Monitor B850 B 4 4 2106778 001 For your notes ...
Страница 190: ...CARESCAPE Monitor B850 C 4 4 2106778 001 For your notes ...
Страница 191: ......