CARESCAPE Monitor B850
116
2106778-001
2.
Select
Information >
Device Information
.
3.
Verify that the device information is correct:
The connected parameter modules are identified.
The connected CARESCAPE Network ID interfaces are identified.
The connected USB input devices are identified.
4.
Stay connected to Webmin.
8.3.4 Configuration Information
1.
Select
Information >
Configuration Information
.
2.
Verify that the patient monitor is correctly configured:
The host Information is correct.
The active software package is correct.
The correct host licenses are enabled.
The correct PDM licenses are enabled.
Patient ID prefix is correctly configured.
Unit and Bed name are correctly configured.
S/5 printers are correctly configured.
IX printers and printer locations are correctly configured.
Remote service is correctly configured.
National requirements are correctly configured.
Network is correctly configured.
Power Line Frequency is correctly configured.
MUSE/12SL is correctly configured.
Citrix is correctly configured.
3.
Log out Webmin.
8.3.5 Trim knob, keypad, and remote
Perform the applicable steps for all the connected displays and for the optional remote
controller.
1.
Press any hard key in the keypad or remote and verify that the selected menu is opened
on the screen or the selected activity is started.
2.
Rotate the
Trim Knob
control in either direction to move from option to option on the
display until you have an active parameter window or main menu item highlighted. Press
the
Trim Knob
control once to select the highlighted option. Verify that the selected
menu is opened on the screen or the selected activity is started.
3.
Select
to close any open menu and return to the main display.
8.3.6 Mouse
Perform this test only if a mouse is connected to the patient monitor.
1.
Move the mouse until the pointer (arrow) is over an active parameter window or a main
menu item you wish to select and click the left mouse button once to select it.
2.
Select
to close the open menu and return to the main display.
Содержание CARESCAPE Monitor B850
Страница 12: ...CARESCAPE Monitor B850 4 2106778 001 For your notes ...
Страница 16: ...CARESCAPE Monitor B850 8 2106778 001 For your notes ...
Страница 56: ...CARESCAPE Monitor B850 48 2106778 001 For your notes ...
Страница 76: ...CARESCAPE Monitor B850 68 2106778 001 For your notes ...
Страница 112: ...CARESCAPE Monitor B850 104 2106778 001 For your notes ...
Страница 128: ...CARESCAPE Monitor B850 120 2106778 001 For your notes ...
Страница 156: ...CARESCAPE Monitor B850 148 2106778 001 ...
Страница 157: ...Troubleshooting 149 2106778 001 Problem No waveforms or parameters are displayed at the CARESCAPE Cental Station ...
Страница 160: ...CARESCAPE Monitor B850 152 2106778 001 For your notes ...
Страница 177: ...Service parts 169 2106778 001 13 3 Exploded view ...
Страница 178: ...CARESCAPE Monitor B850 170 2106778 001 For your notes ...
Страница 182: ...CARESCAPE Monitor B850 A 4 4 2106778 001 Notes Used service parts Signature Date ...
Страница 186: ...CARESCAPE Monitor B850 B 4 4 2106778 001 For your notes ...
Страница 190: ...CARESCAPE Monitor B850 C 4 4 2106778 001 For your notes ...
Страница 191: ......