Hardware installation
53
2106778-001
6
Hardware installation
CAUTION
LOSS OF MONITORING - Leave space for circulation of air to prevent the
monitor from overheating. The manufacturer is not responsible for damage to
equipment caused by improperly vented cabinets, improper or faulty power, or
insufficient wall strength to support equipment mounted on such walls.
WARNING
The parameter modules are not able to withstand unpacked drops from a
height of 1 m without damage. If a module is dropped, please service it
before taking it back into use.
WARNING
After transferring or reinstalling the monitor, always check that it is
properly connected and all parts are securely attached.
WARNING
SITE REQUIREMENTS - Do not route cables or tubing in a way that they may
present a stumbling hazard.
WARNING
EXPLOSION - Do not use this equipment in the presence of flammable
anesthetics, vapors or liquids.
WARNING
EXCESSIVE TOUCH CURRENT - To avoid excessive patient leakage current,
do not simultaneously touch the patient and the electrical connectors
located at the rear panel of the CPU unit or within the module frame.
6.1 Installing a battery into the PDM module
The battery for the PDM module, if included, is shipped separately and need to be installed and
fully charged prior to taking into use.
WARNING
PHYSICAL INJURY- Make sure the battery is completely inserted and the
battery door is completely closed. Falling batteries could seriously or fatally
injure neonatal or other vulnerable patients.
WARNING
EXPLOSION OR FIRE -Using non-recommended batteries could result in
injury/burns to the patients or users. Only use batteries recommended or
manufactured by GE.
WARNING
EXPLOSION HAZARD - Do not incinerate a battery or store at high
temperatures. Serious injury or death could result.
NOTE: Refer to the Module Frames and Modules Service Manual for information about PDM
battery maintenance.
6.1.1 Testing the battery charge
Before installing a battery, verify the battery’s state of charge. Press the green
TEST
button on
the battery. The number of charge level indicator LEDs that illuminate indicates the
approximate charge remaining in the battery.
Four LEDs illuminated: 75% – 100% of full-charge capacity.
Three LEDs illuminated: 50% – 74.9% of full-charge capacity.
Two LEDs illuminated: 25% – 49.9% of full-charge capacity.
One LED illuminated: 10% – 24.9% of full-charge capacity.
Содержание CARESCAPE Monitor B850
Страница 12: ...CARESCAPE Monitor B850 4 2106778 001 For your notes ...
Страница 16: ...CARESCAPE Monitor B850 8 2106778 001 For your notes ...
Страница 56: ...CARESCAPE Monitor B850 48 2106778 001 For your notes ...
Страница 76: ...CARESCAPE Monitor B850 68 2106778 001 For your notes ...
Страница 112: ...CARESCAPE Monitor B850 104 2106778 001 For your notes ...
Страница 128: ...CARESCAPE Monitor B850 120 2106778 001 For your notes ...
Страница 156: ...CARESCAPE Monitor B850 148 2106778 001 ...
Страница 157: ...Troubleshooting 149 2106778 001 Problem No waveforms or parameters are displayed at the CARESCAPE Cental Station ...
Страница 160: ...CARESCAPE Monitor B850 152 2106778 001 For your notes ...
Страница 177: ...Service parts 169 2106778 001 13 3 Exploded view ...
Страница 178: ...CARESCAPE Monitor B850 170 2106778 001 For your notes ...
Страница 182: ...CARESCAPE Monitor B850 A 4 4 2106778 001 Notes Used service parts Signature Date ...
Страница 186: ...CARESCAPE Monitor B850 B 4 4 2106778 001 For your notes ...
Страница 190: ...CARESCAPE Monitor B850 C 4 4 2106778 001 For your notes ...
Страница 191: ......