Installation checkout
111
2106778-001
9.
Read and record the current leakage indicated on the safety tester.
10. Configure the safety analyzer as follows (SFC):
Polarity: REVERSED
Neutral: OPEN
Earth (GND): CLOSED
11. Read and record the current leakage indicated on the safety tester.
12. Configure the safety analyzer as follows (NC):
Polarity: REVERSED
Neutral: CLOSED
Earth (GND): CLOSED
13. Read and record the current leakage indicated on the safety tester.
14. Power off the device under test.
Acceptance criteria in Normal Condition (NC):
•
All readings shall be less than or equal to 100
A
Acceptance criteria in Single Fault Condition (SFC) – earth open or one of the supply conductors
open at a time:
•
All readings shall be less than or equal to 300
A f
or installations that require compliance
to UL 60601-1 requirements.
•
All readings shall be less than or equal to 500
A f
or installations that require compliance
to EN 60601-1 / IEC 60601-1 requirements.
8.2.7 Patient leakage current tests – overview
The following table specifies the parameter modules and the related patient connectors to be
tested in the
8.2.8. Patient (source) leakage current tests
and in the
Use the safety test body kit, P/N M1155870 (or equivalent), to perform patient leakage current
tests. This safety test body kit contains various patient connectors where all pins are shorted
out together. For information on which test body to use for each patient connector, refer to the
service instructions included in the safety test body kit.
NOTE: If not otherwise stated in the table below, each test body is connected directly to the
specified connector in the patient module.
Table 1
Patient connectors to be tested with each module
Module
Patient connector
E-PSMP
ECG &
SpO2
E-PT
P3/P7
E-PP
P5
Содержание CARESCAPE Monitor B850
Страница 12: ...CARESCAPE Monitor B850 4 2106778 001 For your notes ...
Страница 16: ...CARESCAPE Monitor B850 8 2106778 001 For your notes ...
Страница 56: ...CARESCAPE Monitor B850 48 2106778 001 For your notes ...
Страница 76: ...CARESCAPE Monitor B850 68 2106778 001 For your notes ...
Страница 112: ...CARESCAPE Monitor B850 104 2106778 001 For your notes ...
Страница 128: ...CARESCAPE Monitor B850 120 2106778 001 For your notes ...
Страница 156: ...CARESCAPE Monitor B850 148 2106778 001 ...
Страница 157: ...Troubleshooting 149 2106778 001 Problem No waveforms or parameters are displayed at the CARESCAPE Cental Station ...
Страница 160: ...CARESCAPE Monitor B850 152 2106778 001 For your notes ...
Страница 177: ...Service parts 169 2106778 001 13 3 Exploded view ...
Страница 178: ...CARESCAPE Monitor B850 170 2106778 001 For your notes ...
Страница 182: ...CARESCAPE Monitor B850 A 4 4 2106778 001 Notes Used service parts Signature Date ...
Страница 186: ...CARESCAPE Monitor B850 B 4 4 2106778 001 For your notes ...
Страница 190: ...CARESCAPE Monitor B850 C 4 4 2106778 001 For your notes ...
Страница 191: ......