Using Webmin service interface
45
2106778-001
Webmin module
Use the module
Admit Settings
to configure Patient ID Prefix and barcode settings.
Certificate Management
to manage SSL certificates installed on the system. With
this module, self-signed certificates and certificate
signing requests (CSR) can be generated. Certificates
can be imported, exported, deleted and viewed.
Citrix
to configure the Citrix client settings for the iPanel
software: server address, initial program, Citrix session
timeout, username, password, encryption level.
Host Asset Settings
to enter a host asset number and to view the host serial
number.
Language
to select the language used in clinical user interface and
to select the keyboard locale setting for the
alphanumeric keyboard and the barcode reader.
Licenses
to enable and activate a software package, to enable
and activate software features and to upload and
activate a license file.
MUSE/12SL
to configure the host for sending and viewing 12SL
information.
Modules
to configure some acquisition module specific settings
after corrective maintenance, or for administrative
purposes. These settings are saved to the permanent
memory of the related acquisition module and the
settings travel with the module from one patient monitor
to another.
Refer to the Module Frames and Modules Service
Manual for detailed information how to change these
settings.
National Requirements
to activate France specific defaults for the ECG HR
adjustment range and the reminder beep behavior.
Network
to select and configure the real-time network as the MC
Network (CARESCAPE Network or S/5 Network). It also
allows the user to configure the IX and ADT Network
settings.
Passwords
to change the passwords for the biomed and clinical
users.
Power frequency
to set the power line frequency.
Printers
to configure the patient monitor to print to up to 12 laser
printers connected on the IX Network. There are
sub-modules for installing a printer, deleting a printer
and for printing a test page.
Remote Service
to configure and control the InSite RSvP remote service
tool.
Содержание CARESCAPE Monitor B850
Страница 12: ...CARESCAPE Monitor B850 4 2106778 001 For your notes ...
Страница 16: ...CARESCAPE Monitor B850 8 2106778 001 For your notes ...
Страница 56: ...CARESCAPE Monitor B850 48 2106778 001 For your notes ...
Страница 76: ...CARESCAPE Monitor B850 68 2106778 001 For your notes ...
Страница 112: ...CARESCAPE Monitor B850 104 2106778 001 For your notes ...
Страница 128: ...CARESCAPE Monitor B850 120 2106778 001 For your notes ...
Страница 156: ...CARESCAPE Monitor B850 148 2106778 001 ...
Страница 157: ...Troubleshooting 149 2106778 001 Problem No waveforms or parameters are displayed at the CARESCAPE Cental Station ...
Страница 160: ...CARESCAPE Monitor B850 152 2106778 001 For your notes ...
Страница 177: ...Service parts 169 2106778 001 13 3 Exploded view ...
Страница 178: ...CARESCAPE Monitor B850 170 2106778 001 For your notes ...
Страница 182: ...CARESCAPE Monitor B850 A 4 4 2106778 001 Notes Used service parts Signature Date ...
Страница 186: ...CARESCAPE Monitor B850 B 4 4 2106778 001 For your notes ...
Страница 190: ...CARESCAPE Monitor B850 C 4 4 2106778 001 For your notes ...
Страница 191: ......