CARESCAPE Monitor B850
70
2106778-001
4.
Select
Save
.
The hostname configuration will take effect immediately.
7.2.2 Selecting and configuring CARESCAPE Network
To select CARESCAPE Network MC for the real-time network infrastructure:
1.
Log in to Webmin.
2.
Select
Configuration >
Network > Wired Interfaces
.
The current network configuration is shown in the
Present Configuration
table.
3.
Below
Select Network Type
in the
Network Configuration
window, select the applicable
network type:
CARESCAPE Network
.
4.
Select
Next
to proceed on network configuration.
NOTE: When
CARESCAPE Network
is selected, the S/5 Network is disabled.
The patient monitor may be configured to the MC Network, or IX Network, or both.
NOTE: The following IP addresses are reserved for the monitor’s internal use: 192.168.249.x,
192.168.250.x, 192.168.252.x, 192.168.253.x, 192.168.254.1 and 192.168.254.2.
5.
In the
MC Network
area
.
a.
Enter a
Static IP
address.
b.
Enter a valid
Netmask
level.
c.
Select
Speed & Duplex
option.
6.
If separate MC Network segments are connected together by a router, configure the
router information below the MC static route:
a.
Enter
Destination Address
.
b.
Enter
Destination Netmask
.
c.
Enter an
MC Gateway.
7.
In the
IX Network
area, select
DHCP
or
Manual Configuration
.
If
Manual Configuration
is selected, enter the following information in the entry fields:
a.
Enter a
Static IP
address.
b.
Enter a valid
Netmask
level.
c.
Enter a valid
Default Gateway
.
d.
Enter valid
DNS Server
addresses if applicable.
8.
Enter a unique
Custom IX MAC Address.
By default, the MC Network MAC address is
shown here. Make sure that you do not use a MAC address that is already in use on your
network.
NOTE: The following addresses are reserved for the monitor’s internal use (RFC 7042):
•
00:00:5E:xx:xx:xx: Reserved for IANA unicast
•
01:00:5E:xx:xx:xx: Reserved for IANA multicast
•
33:33:xx:xx:xx:xx: Reserved for IPv6 multicast
Network connection(s)
Steps to be performed
MC Network only
5, (6), and 10
IX Network only
7, 8, 9 and 10
Both MC and IX Network
5, (6), 7, 8, 9 and 10
Содержание CARESCAPE Monitor B850
Страница 12: ...CARESCAPE Monitor B850 4 2106778 001 For your notes ...
Страница 16: ...CARESCAPE Monitor B850 8 2106778 001 For your notes ...
Страница 56: ...CARESCAPE Monitor B850 48 2106778 001 For your notes ...
Страница 76: ...CARESCAPE Monitor B850 68 2106778 001 For your notes ...
Страница 112: ...CARESCAPE Monitor B850 104 2106778 001 For your notes ...
Страница 128: ...CARESCAPE Monitor B850 120 2106778 001 For your notes ...
Страница 156: ...CARESCAPE Monitor B850 148 2106778 001 ...
Страница 157: ...Troubleshooting 149 2106778 001 Problem No waveforms or parameters are displayed at the CARESCAPE Cental Station ...
Страница 160: ...CARESCAPE Monitor B850 152 2106778 001 For your notes ...
Страница 177: ...Service parts 169 2106778 001 13 3 Exploded view ...
Страница 178: ...CARESCAPE Monitor B850 170 2106778 001 For your notes ...
Страница 182: ...CARESCAPE Monitor B850 A 4 4 2106778 001 Notes Used service parts Signature Date ...
Страница 186: ...CARESCAPE Monitor B850 B 4 4 2106778 001 For your notes ...
Страница 190: ...CARESCAPE Monitor B850 C 4 4 2106778 001 For your notes ...
Страница 191: ......