CARESCAPE Monitor B850
54
2106778-001
One LED flashing: < 10% of full-charge capacity remaining.
NOTE: Keep the PDM connected to the patient monitor and the patient monitor connected to
the AC mains until the PDM battery is fully charged.
6.1.2 Installing the battery into the PDM module
1.
Open the battery door by gently pulling on the battery door pull tab.
2.
Pull the battery tray out of the PDM using the battery tray strap.
3.
Insert the battery with the
TEST
button facing up and the arrow pointing into the PDM.
4.
Press the battery door closed until it seals the battery compartment.
6.2 Mounting the patient monitor
The patient monitor has an integrated GCX mounting plate. This facilitates all mounting options
for the patient monitor. Refer to the CARESCAPE Modular Monitors Mounting Solutions to
identify the compatible mounting hardware for the patient monitor.
6.2.1 Installing the CPU
WARNING
EQUIPMENT MALFUNCTION - Using the CPU in a wrong position may result
in equipment malfunction. The CPU must only be used in vertical or
horizontal position. To avoid the risk of equipment malfunction, make sure
that the CPU is not used in any other position.
WARNING
EQUIPMENT MALFUNCTION - When the CPU is used in vertical position, it is
not protected against harmful effects of dripping water (IPX1). Therefore,
Содержание CARESCAPE Monitor B850
Страница 12: ...CARESCAPE Monitor B850 4 2106778 001 For your notes ...
Страница 16: ...CARESCAPE Monitor B850 8 2106778 001 For your notes ...
Страница 56: ...CARESCAPE Monitor B850 48 2106778 001 For your notes ...
Страница 76: ...CARESCAPE Monitor B850 68 2106778 001 For your notes ...
Страница 112: ...CARESCAPE Monitor B850 104 2106778 001 For your notes ...
Страница 128: ...CARESCAPE Monitor B850 120 2106778 001 For your notes ...
Страница 156: ...CARESCAPE Monitor B850 148 2106778 001 ...
Страница 157: ...Troubleshooting 149 2106778 001 Problem No waveforms or parameters are displayed at the CARESCAPE Cental Station ...
Страница 160: ...CARESCAPE Monitor B850 152 2106778 001 For your notes ...
Страница 177: ...Service parts 169 2106778 001 13 3 Exploded view ...
Страница 178: ...CARESCAPE Monitor B850 170 2106778 001 For your notes ...
Страница 182: ...CARESCAPE Monitor B850 A 4 4 2106778 001 Notes Used service parts Signature Date ...
Страница 186: ...CARESCAPE Monitor B850 B 4 4 2106778 001 For your notes ...
Страница 190: ...CARESCAPE Monitor B850 C 4 4 2106778 001 For your notes ...
Страница 191: ......