Hardware installation
55
2106778-001
we recommend using the CPU Flush Mount Kit with environment shield to
provide protection from drips or splashes of liquids. Without the shield,
liquids may enter the equipment and lead to its malfunction.
For vertical mounting configurations, a flush mount is available. The flush mount lowers the
probability of the ingress of fluids into the assembly. The mount can be installed in two
orientations and should be placed to ensure the top surface of the patient monitor is protected.
For details, refer to the installation instructions included with the flush mount.
NOTE: The GE logo can be rotated for vertical mounting configurations.
WARNING
Never install equipment above the patient.
WARNING
Use only manufacturer specified mounts.
Содержание CARESCAPE Monitor B850
Страница 12: ...CARESCAPE Monitor B850 4 2106778 001 For your notes ...
Страница 16: ...CARESCAPE Monitor B850 8 2106778 001 For your notes ...
Страница 56: ...CARESCAPE Monitor B850 48 2106778 001 For your notes ...
Страница 76: ...CARESCAPE Monitor B850 68 2106778 001 For your notes ...
Страница 112: ...CARESCAPE Monitor B850 104 2106778 001 For your notes ...
Страница 128: ...CARESCAPE Monitor B850 120 2106778 001 For your notes ...
Страница 156: ...CARESCAPE Monitor B850 148 2106778 001 ...
Страница 157: ...Troubleshooting 149 2106778 001 Problem No waveforms or parameters are displayed at the CARESCAPE Cental Station ...
Страница 160: ...CARESCAPE Monitor B850 152 2106778 001 For your notes ...
Страница 177: ...Service parts 169 2106778 001 13 3 Exploded view ...
Страница 178: ...CARESCAPE Monitor B850 170 2106778 001 For your notes ...
Страница 182: ...CARESCAPE Monitor B850 A 4 4 2106778 001 Notes Used service parts Signature Date ...
Страница 186: ...CARESCAPE Monitor B850 B 4 4 2106778 001 For your notes ...
Страница 190: ...CARESCAPE Monitor B850 C 4 4 2106778 001 For your notes ...
Страница 191: ......