CARESCAPE Monitor B850
126
2106778-001
Planned maintenance schedule
Perform the planned maintenance procedure every 2 years after installation. Perform the
procedure in the following order:
1.
Visual inspection (section 10.1)
2.
Electrical safety tests (section 10.2)
3.
Functional check (section 10.3, all steps)
Replace the CPU timekeeper battery every 5 years, or whenever the
Service Monitor Error
Code 0xHOST1100
message is shown.
10.1 Visual inspection
Follow the procedure in section
10.2 Electrical safety checks
Electrical safety tests provide a method of determining if potential electrical health hazards to
the patient or operator of the device exist.
Perform the following electrical safety tests described in detail in chapter
:
8.2.5. Ground leakage current test
Record the values of the tests on the Appendix
.
10.3 Functional check
10.3.1 Start-up
Follow the procedure in section
.
10.3.2 Display
Follow the procedure in section
10.3.3
PSM / PDM identification
1.
Configure the ECG1 waveform field and the NIBP parameter window to the patient
monitor screen with adequate priority. Connect a PSM or PDM module to the patient
monitor.
2.
Verify that the ECG waveform field, the NIBP parameter window and the related
information appear on the patient monitor screen.
10.3.4
E-module identification
1.
Log in to Webmin.
Содержание CARESCAPE Monitor B850
Страница 12: ...CARESCAPE Monitor B850 4 2106778 001 For your notes ...
Страница 16: ...CARESCAPE Monitor B850 8 2106778 001 For your notes ...
Страница 56: ...CARESCAPE Monitor B850 48 2106778 001 For your notes ...
Страница 76: ...CARESCAPE Monitor B850 68 2106778 001 For your notes ...
Страница 112: ...CARESCAPE Monitor B850 104 2106778 001 For your notes ...
Страница 128: ...CARESCAPE Monitor B850 120 2106778 001 For your notes ...
Страница 156: ...CARESCAPE Monitor B850 148 2106778 001 ...
Страница 157: ...Troubleshooting 149 2106778 001 Problem No waveforms or parameters are displayed at the CARESCAPE Cental Station ...
Страница 160: ...CARESCAPE Monitor B850 152 2106778 001 For your notes ...
Страница 177: ...Service parts 169 2106778 001 13 3 Exploded view ...
Страница 178: ...CARESCAPE Monitor B850 170 2106778 001 For your notes ...
Страница 182: ...CARESCAPE Monitor B850 A 4 4 2106778 001 Notes Used service parts Signature Date ...
Страница 186: ...CARESCAPE Monitor B850 B 4 4 2106778 001 For your notes ...
Страница 190: ...CARESCAPE Monitor B850 C 4 4 2106778 001 For your notes ...
Страница 191: ......