Installation checkout
117
2106778-001
8.3.7 Alphanumeric keyboard
Perform this test only if an alphanumeric keyboard is connected to the patient monitor.
1.
Select
Data & Pages > Admit/Discharge
(or
Start / End Case
).
2.
Select the
Patient tab > Edit Name & MRN
.
3.
Press
Enter
to highlight the
Medical Record Number
field.
4.
Type some text into the
Medical Record Number
field using the connected alphanumeric
keyboard. Include some characters that are specific to the chosen keyboard locale to
verify that the keyboard language configuration is correct.
5.
Select
to close the open menu and return to the main display.
8.3.8 Barcode reader
Perform this test only if a barcode reader is connected to the patient monitor.
1.
Select
Data & Pages > Admit/Discharge
(or
Start / End Case
).
2.
Select the
Patient
tab.
3.
Scan a test barcode that is applicable to your system:
8.3.9 MC Network and S/5 Network
Perform the following test only if the patient monitor is connected to a wired MC Network or S/5
Network.
NOTE: Make sure that at least one other patient monitor is on the network. The other patient
monitor must be in an admitted state and have an active ECG measurement with a simulator
signal.
Parser Type
Test Procedure
Length Delimited or Character
Delimited Parser
1.
2.
3.
4. Select
Scan from Barcode
.
5. Scan a known test barcode obtained from the hospital.
NOTE: The barcode data content must be known and in
compliance with the completed parser configuration.
6. Verify that the data content in the barcode is correctly
populated to the related fields in the
Patient
and the
Administr. Information
tabs.
No parser
1.
2.
3.
4. Select
Edit Name & MRN
and press
Enter
to highlight
the
Medical Record Number
field.
5. Scan a sample barcode that only contains one piece of
information (e.g., a Serial Number barcode from a
module’s device label).
6. Verify that the data is correctly populated into the
Medical Record Number
field.
7. Select
to close the open menu and return to the main display.
Содержание CARESCAPE Monitor B850
Страница 12: ...CARESCAPE Monitor B850 4 2106778 001 For your notes ...
Страница 16: ...CARESCAPE Monitor B850 8 2106778 001 For your notes ...
Страница 56: ...CARESCAPE Monitor B850 48 2106778 001 For your notes ...
Страница 76: ...CARESCAPE Monitor B850 68 2106778 001 For your notes ...
Страница 112: ...CARESCAPE Monitor B850 104 2106778 001 For your notes ...
Страница 128: ...CARESCAPE Monitor B850 120 2106778 001 For your notes ...
Страница 156: ...CARESCAPE Monitor B850 148 2106778 001 ...
Страница 157: ...Troubleshooting 149 2106778 001 Problem No waveforms or parameters are displayed at the CARESCAPE Cental Station ...
Страница 160: ...CARESCAPE Monitor B850 152 2106778 001 For your notes ...
Страница 177: ...Service parts 169 2106778 001 13 3 Exploded view ...
Страница 178: ...CARESCAPE Monitor B850 170 2106778 001 For your notes ...
Страница 182: ...CARESCAPE Monitor B850 A 4 4 2106778 001 Notes Used service parts Signature Date ...
Страница 186: ...CARESCAPE Monitor B850 B 4 4 2106778 001 For your notes ...
Страница 190: ...CARESCAPE Monitor B850 C 4 4 2106778 001 For your notes ...
Страница 191: ......