Maintenance and checkout
125
2106778-001
10 Maintenance and checkout
This chapter specifies the checkout procedure and the maintenance activities to be performed
to the patient monitor after corrective maintenance and during planned maintenance.
WARNING
Only perform maintenance procedures specifically described in the manual.
WARNING
Planned maintenance should be carried out at recommended interval.
Failure to implement the recommended maintenance schedule may cause
equipment failure and possible health hazards.
NOTE: The manufacturer does not, in any manner, assume the responsibility for performing the
recommended maintenance schedule, unless an Equipment Maintenance Agreement exists.
The sole responsibility rests with the individuals, hospitals, or institutions utilizing the device.
NOTE: Refer to Module Frames and Modules Service Manual for corrective and planned
maintenance checkout procedures of the parameter modules.
NOTE: Refer to the PDM section in the Module Frames and Modules Service Manual for the
battery maintenance procedure of the PDM battery.
Corrective maintenance
Service personnel shall perform the following checkout procedure steps after any corrective
maintenance, before taking the monitor back into clinical use:
Performed service activity
Required checkout procedure
Visual
inspections
(section 10.1)
Electrical
safety test
(section 10.2)
Functional check
(section 10.3)
After detaching or replacing:
- Power supply
- CPU PCB assembly
- Lithium -ion battery (timekeeper)
Yes
Yes
- All steps
After detaching or replacing:
- 3rd video card
Yes
Yes
- 10.3.1 Start-up
- 10.3.2 Display
- 10.3.3 PSM / PDM
After detaching or replacing:
- Enclosure Kit
- Hardware kit (or any part of this)
- Speaker assembly
Yes
Yes
After detaching or replacing:
- Label kit
Yes
No
After detaching or replacing:
- uDOM
Yes
Yes
- All steps
After detaching or replacing:
- M-Port keypad Kit
- Remote Control Kit
Yes
No
-
-
Содержание CARESCAPE Monitor B850
Страница 12: ...CARESCAPE Monitor B850 4 2106778 001 For your notes ...
Страница 16: ...CARESCAPE Monitor B850 8 2106778 001 For your notes ...
Страница 56: ...CARESCAPE Monitor B850 48 2106778 001 For your notes ...
Страница 76: ...CARESCAPE Monitor B850 68 2106778 001 For your notes ...
Страница 112: ...CARESCAPE Monitor B850 104 2106778 001 For your notes ...
Страница 128: ...CARESCAPE Monitor B850 120 2106778 001 For your notes ...
Страница 156: ...CARESCAPE Monitor B850 148 2106778 001 ...
Страница 157: ...Troubleshooting 149 2106778 001 Problem No waveforms or parameters are displayed at the CARESCAPE Cental Station ...
Страница 160: ...CARESCAPE Monitor B850 152 2106778 001 For your notes ...
Страница 177: ...Service parts 169 2106778 001 13 3 Exploded view ...
Страница 178: ...CARESCAPE Monitor B850 170 2106778 001 For your notes ...
Страница 182: ...CARESCAPE Monitor B850 A 4 4 2106778 001 Notes Used service parts Signature Date ...
Страница 186: ...CARESCAPE Monitor B850 B 4 4 2106778 001 For your notes ...
Страница 190: ...CARESCAPE Monitor B850 C 4 4 2106778 001 For your notes ...
Страница 191: ......