Test Setup - Test Configurator, Timer, and System
292
88000 Series
RFC 2544 - Global
Coupled
is available with
Bidirectional
flow direction. When the
Coupled
check box is:
Cleared, the decision to increase or decrease the TX Rate of the
next iteration is taken independently per direction.
Selected, the decision to increase or decrease the TX Rate of the
next iteration is taken commonly for both directions.
Rate Unit
determines the unit used to display the rate values:
%
,
Mbit/s
, or
Gbit/s
.
Pass/Fail Verdict
check box when selected (default) enables the use
of the pass/fail verdict.
Subtests and Estimated Time
Subtests
allows enabling the
Throughput
,
Back-to-Back
,
Frame Loss
,
and
Latency
subtests individually.
Estimated Time (H:MM)
indicates the estimated time required to
complete each subtest at best conditions. The total estimated time to
complete all subtests is also displayed.
Frame Distribution
Frame Distribution
allows selecting either
RFC 2544
(default) or
User
Defined
distribution.
Quantity
is only available when
User Defined
is selected and allows
selecting the number of frames, from
1
to
10
(
7
by default), in the
distribution.
Содержание Power Blazer 88200NGE
Страница 1: ...www EXFO com User Guide 88000 Series Power Blazer HIGH SPEED MULTISERVICE TEST MODULE...
Страница 14: ......
Страница 20: ......
Страница 64: ...Test Setup Test Applications 54 88000 Series OTN BERT...
Страница 65: ...Test Setup Test Applications Power Blazer 55 OTN BERT OTU4 for 88200NGE 890 890NGE OTU4 for 88260...
Страница 71: ...Test Setup Test Applications Power Blazer 61 SONET SDH BERT For SDH BERT...
Страница 76: ...Test Setup Test Applications 66 88000 Series SONET SDH DSn PDH BERT For SDH...
Страница 620: ......
Страница 726: ......
Страница 752: ......
Страница 756: ......
Страница 768: ......
Страница 789: ...Glossary Power Blazer 779 10G Ethernet Client The following figure presents a typical network application...
Страница 803: ...Glossary Power Blazer 793 Ethernet Cables Straight Through Cable 1000 Mbit s Crossover Cable 1000 Mbit s...
Страница 847: ...Glossary Power Blazer 837 SONET DSn SDH PDH The GUI Grid indicates the TUG 2 x and TUG 3 x values...
Страница 886: ...0 5 1 5 48 5 0 176 W d y K y K D...