Test Setup - Test Configurator, Timer, and System
Power Blazer
187
EtherBERT, FC BERT, BERT (eCPRI, CPRI/OBSAI, and FlexE), and Unframed BERT
FC Frame
Note:
Only available with FC BERT.
Frame Size (Bytes)
allows entering the frame size (in step of 4 bytes) for
FC BERT test application:
40
to
2148
bytes (default). However the
minimum frame size is 64 bytes when the
Latency Tag
check box is
selected.
Latency Tag Insertion
Note:
Only available with FC BERT.
Round trip latency measures the average round trip latency (delay)
between the time the data was transmitted and received.
Latency Tag
check box when selected enables latency
measurements.
Pass/Fail Verdict
check box when selected enables the pass/fail
verdict and allows setting the
Round-Trip Latency Threshold
.
Round-Trip Latency Threshold (ms)
allows the selection of the round
trip latency:
0.015
(default) to
8000 ms
.
Содержание Power Blazer 88200NGE
Страница 1: ...www EXFO com User Guide 88000 Series Power Blazer HIGH SPEED MULTISERVICE TEST MODULE...
Страница 14: ......
Страница 20: ......
Страница 64: ...Test Setup Test Applications 54 88000 Series OTN BERT...
Страница 65: ...Test Setup Test Applications Power Blazer 55 OTN BERT OTU4 for 88200NGE 890 890NGE OTU4 for 88260...
Страница 71: ...Test Setup Test Applications Power Blazer 61 SONET SDH BERT For SDH BERT...
Страница 76: ...Test Setup Test Applications 66 88000 Series SONET SDH DSn PDH BERT For SDH...
Страница 620: ......
Страница 726: ......
Страница 752: ......
Страница 756: ......
Страница 768: ......
Страница 789: ...Glossary Power Blazer 779 10G Ethernet Client The following figure presents a typical network application...
Страница 803: ...Glossary Power Blazer 793 Ethernet Cables Straight Through Cable 1000 Mbit s Crossover Cable 1000 Mbit s...
Страница 847: ...Glossary Power Blazer 837 SONET DSn SDH PDH The GUI Grid indicates the TUG 2 x and TUG 3 x values...
Страница 886: ...0 5 1 5 48 5 0 176 W d y K y K D...