Technical data
Proline Promass E TB2 HART
100
Hauser
18.6
Performance characteristics
Reference operating
conditions
• Error limits following ISO/DIS 11631
• Water with +20 to +30
°
C (+68 to +86
°
F) at 2 to 4 bar (29 to 58 psi)
• Specifications as per calibration protocol ±5
°
C (±9
°
F) and ±2 bar (±29 psi)
• Accuracy based on accredited calibration rigs that are traced to ISO 17025
To obtain measured errors, use the
Applicator
Maximum measured error
In addition to the values indicated, the measured error at the current output is typically ±4
m
A.
o.r. = of reading; 1 g/cm
3
= 1 kg/l; T = fluid temperature
Mass flow and volume flow (liquids)
±0.25 % o.r.
Mass flow (gases)
±0.75 % o.r.
Density (liquids)
• ±0.0005 g/cm
3
(under reference conditions)
• ±0.0005 g/cm
3
(after field density calibration under process conditions)
• ±0.02 g/cm
3
(over the entire measuring range of the sensor)
Temperature
±
0.5
°
C
±
0.005
×
T °C (±0.9
°
F
±
0.005
×
(T – 32) °F)
Zero point stability
DN
Zero point stability
[mm]
[in]
[kg/h] or [l/h]
[lb/min]
8
3/8
0.24
0.0088
15
½
0.78
0.0287
25
1
2.16
0.0794
40
1½
5.40
0.1985
50
2
8.40
0.3087