10.4.6 Configuring the partial filled pipe detection
The
Partial filled pipe detection
wizard guides you systematically through all parameters that
have to be set for configuring the monitoring of the pipe filling.
Navigation path
"Setup" menu
®
Partial filled pipe detection
Sequence of the wizard
Assign process
variable
Off
Mass flow
Volume flow
Corrected
volume flow
High value
partial filled
pipe detection
Low value
partial filled
pipe detection
Response time
part. filled
pipe detect.
End of wizard
A0013801-EN
å
17
"Partial filled pipe detection" wizard in the "Setup" menu
Parameter overview with brief description
Parameters
Description
Selection/
User entry
Factory
setting
Assign process variable
Select a process variable to detect
empty or partially filled pipes.
• Off
• Density
• Reference density
Density
Low value partial filled pipe
detection
Enter a lower limit value to
activate detection of an empty or
partially filled pipe.
Max. 15-digit, positive floating-
point number
Country-
dependent:
• 0.2 kg/l
• 12.5 lb/cf
High value partial filled pipe
detection
Enter an upper limit value to
activate detection of an empty or
partially filled pipe.
Max. 15-digit, positive floating-
point number
Country-
dependent:
• 6 kg/l
• 374.6 lb/cf
Response time part. filled pipe
detect.
Enter the time interval until the
diagnostics message S862
Partly filled pipe detection
is
displayed for an empty or partially
filled pipe.
0 to 100 s
1 s