Diagnostics and troubleshooting
Proline Promass E TB2 HART
86
Hauser
Diagnostics event
Event text
Remedial measures
Event level from the
factory
S044
Sensor drift
Electrodynamic sensor signal deviation!
1. Check or change main electronics.
2. Change sensor.
Warning
S046
Sensor limit
The measuring tube oscillation frequency is
outside the permitted range!
1. Inspect sensor.
2. Check process condition.
Warning
F062
Sensor connection
1. Change main electronic module.
2. Change sensor.
Alarm
F082
Data storage
1. Change main electronic module.
2. Change sensor.
Alarm
F083
Memory content
1. Restart device.
2. Restore S-Dat data.
3. Change sensor.
Alarm
1)
Event level can be changed: Chapter 12.5.1 "Adapting event level"
12.5.3 Diagnostics for the electronics
Diagnostics event
Event text
Remedial measures
Event level from the
factory
F242
Software incompatible 1. Check software.
2. Flash or change main electronics module.
Alarm
F252
Modules incompatible 1. Check electronic modules.
2. Change I/O or main electronics module.
Alarm
F261
Electr. modules
1. Restart device.
2. Check electronic modules.
3. Change I/O Modul or main electronics.
Alarm
F262
Module connect.
1. Check module connections.
2. Change electronics modules or sensor.
Alarm
F270
Main electronic
Change main electronics module.
Alarm
F271
Main electronic
1. Restart device.
2. Change main electronic module.
Alarm
F272
Main electronic
1. Restart device.
2. Contact service.
Alarm
F273
Main electronics
Emergency operation via display!
Change main electronics.
Alarm
S274
Main electronic
Unstable measurement!
Change the main electronics.
Warning
F275
I/O module
Change I/O module.
Alarm
F276
I/O module
1. Restart device.
2. Change I/O module.
Alarm
F282
Data storage
1. Restart device.
2. Contact service.
Alarm
F283
Memory content
1. Transfer data or reset device.
2. Contact service.
Alarm
F311
Electronics error
1. Transfer data or reset device.
2. Contact service.
Alarm
M311
Electronics error
Maintenance required!
1. Do not perform reset.
2. Contact service.
Warning
1)
Event level can be changed: Chapter 12.5.1 "Adapting event level"