Commissioning
Proline Promass E TB2 HART
68
Hauser
Pressure unit
Parameter overview with brief description
Parameters
Description
Selection/
User entry
Factory setting
Mass flow unit
Select the unit for mass flow.
Result
The selected unit applies for:
– Current outputs
– Low flow cut off
– Simulation process
variable
Unit picklist
Country-dependent:
• kg/h
• lb/min
Mass unit
Select the unit for mass.
Result
The selected unit is taken
from:
Mass flow unit
Unit picklist
Country-dependent:
• kg
• lb
Volume flow unit
Select the unit for volume
flow.
Result
The selected unit applies for:
– Current outputs
– Low flow cut off
– Simulation process
variable
Unit picklist
Country-dependent:
• l/h
• gal/min (us)
Volume
Select the unit for volume.
Result
The selected unit is taken
from:
Volume flow unit
Country-dependent
• l
• gal (us)
Corrected volume flow unit Select the unit for corrected
volume flow.
Result
The selected unit applies for:
– Current outputs
– Low flow cut off
– Simulation process
variable
Unit picklist
Country-dependent:
• Nl/h
• Scf/min
Corrected volume unit
Select the unit for standard
volume.
Result
The selected unit is taken
from:
Corrected volume flow unit
Unit picklist
Country-dependent:
• Nl
• Scf
Density unit
Select the unit for density.
Result
The selected unit applies for:
– Current outputs
– Low value partial filled
pipe detection
– High value partial filled
pipe detection
– Simulation process
variable
– Density adjustment (in the
Expert
menu)
Unit picklist
Country-dependent
• kg/l
• lb/cf