Technical data
Proline Promass E TB2 HART
98
Hauser
g
max(F)
Maximum full scale value for liquid [kg/h]
g
max(G)
<
g
max(F)
g
max(G)
can never be greater than
g
max(F)
r
G
Gas density in [kg/m
3
] at operating conditions
DN
x
[mm]
[in]
[kg/m
3
]
8
3/8
85
15
½
110
25
1
125
40
1 ½
125
50
2
125
To calculate the measuring range, use the
Applicator
Calculation example for gas
• Sensor: Promass E, DN 50
• Gas: Air with a density of 60.3 kg/m
:
(at 20
°
C and 50 bar)
• Measuring range (liquid): 70
000 kg/h
• x = 125 kg/m
3
(for Promass E DN 50)
Maximum possible full scale value:
g
max(G)
=
g
max(F)
×
r
G
:
x = 70
000 kg/h
×
60.3 kg/m
3
:
125 kg/m
:
= 33
800 kg/h
Recommended measuring range
Operable flow range
Greater than 1000: 1. Flow rates above the preset full scale value do not overload the amplifier so
the totalized values are registered correctly.
18.4
Output
Output signal
2-wire:
• Current output 1: 4-20 mA HART
• Current output 2: 4-20 mA
• Lower-range value, full scale value and time constant (0.07 to 999 s) can be set
Signal on alarm
Maintenance information can be viewed via the following interfaces:
• Local display:
– Error symbol (in accordance with NAMUR Recommendation NE 107)
– Plain text display
• Current output: failsafe mode selectable (in accordance with NAMUR Recommendation NE 43):
– Minimum alarm: 3.6 mA
– Maximum alarm (= factory setting): 22 mA
• Digital interface such as HART communication or CDI service interface (in accordance with
NAMUR Recommendation NE 107)
Galvanic isolation
All circuits for the outputs are galvanically isolated from each other.