Proline Promass E TB2 HART
Integrating a device via HART protocol
Hauser
49
9
Integrating a device via HART protocol
9.1
Overview of device description files
Version data for the device
Firmware version
01.00.zz
• On the title page of the Operating Instructions
• On transmitter nameplate (
• Parameter
firmware version
Diagnostics
®
Device info
®
Firmware version
Release date of firmware
version
06.2010
On the title page of the Operating Instructions
Manufacturer ID
0x11
Manufacturer ID
parameter
Diagnostics
®
Device info
®
Manufacturer ID
Device type ID
0x54
Device type
parameter
Diagnostics
®
Device info
®
Device type
HART protocol revision
6.0
---
Device revision
1
•
Device revision
parameter
Diagnostics
®
Device info
®
Device revision
The following is a list of the suitable device description file for each individual operating tool with
source.
Operating tools
Operating tool
Sources for obtaining device descriptions
Field Xpert SFX100
Use update function of handheld terminal
FieldCare
•
®
Download Area
• CD–ROM (contact Hauser)
• DVD (contact Hauser)
AMS Device Manager
(Emerson Process Management)
®
Download Area
SIMATIC PDM
(Siemens)
®
Download Area
Field Communicator 375, 475
(Emerson Process Management)
Use update function of handheld terminal
9.2
Device variables and measured values
The following measured values are assigned to the device variables at the factory:
Device variables
Measured values
Primary device variable (PV)
Mass flow
Secondary device variable (SV)
Totalizer 1
Tertiary device variable (TV)
Density
Quaternary device variable (QV)
Temperature
The assignment of the measured values to the device variables can be modified and assigned as
desired via local operation and the operating tool using the following parameters: