Commissioning
Proline Promass E TB2 HART
64
Hauser
10.4.5 Configuring the low flow cut off
The
Low flow cut off
wizard guides you systematically through all parameters that have to be set
for configuring the low flow cut off.
Navigation path
"Setup" menu
®
Low flow cut off
Sequence of the wizard
Assign process
variable
Off
Mass flow
Volume flow
Corrected volume
flow
Off value
low flow
cutoff
On value
low flow
cutoff
Pressure shock
suppression
End of wizard
A0013799-EN
å
16
"Low flow cut off" wizard in the "Setup" menu
Parameter overview with brief description
Parameters
Description
Selection/
User entry
Factory setting
Assign process variable
Select the process variable for
low flow cut off.
• Off
• Mass flow
• Volume flow
• Corrected volume flow
Mass flow
On value low flow cut off
Enter the on-value for the
low flow cut off.
Max. 15-digit, positive
floating-point number
For liquids: depends on
country and nominal
diameter
Off value low flow cut off
Enter the off-value for the
low flow cut off.
0 to 100
%
50
%
Pressure shock suppression
Enter the time interval for
signal suppression (= active
pressure shock suppression).
0 to 100 s
0 s