Mounting
Proline Promass E TB2 HART
24
Hauser
6.3
Post-mounting check
m
Is the device damaged (visual inspection)?
m
Does the measuring device conform to the measuring point specifications?
For example:
• Process temperature (
• Process pressure (refer to the chapter on "Material load curves" of the "Technical Information" document)
• Ambient temperature range (
m
Has the correct orientation for the sensor been selected (
• According to sensor type
• According to fluid temperature
• According to fluid properties (outgassing, with entrained solids)
m
Does the arrow on the sensor nameplate match the direction of flow of the fluid through the
piping(
m
Are the measuring point identification and labeling correct (visual inspection)?
m
Is the device adequately protected from precipitation and direct sunlight?
m
Are the securing screw and securing clamp tightened securely?