Proline Promass E TB2 HART
Commissioning
Hauser
73
While this action is in progress, the configuration cannot be edited via the local display and a
message on the processing status appears on the display.
10.7
Simulation
The
Simulation
submenu enables you to simulate, without a real flow situation, various process
variables in the process and the device alarm mode and to verify downstream signal chains (switching
valves or closed-control loops).
Navigation path
"Diagnostics" menu
®
Simulation
Structure of the submenu
Simulation
®
Assign simulation process
variable
Value process variable
Simulation current output 1
Value current output 1
Simulation current output 2
Value current output 2
Simulation device alarm
Parameter overview with brief description
Parameters
Prerequisite
Description
Selection/
User entry
Factory setting
Assign simulation
process variable
-
Select a process
variable for the
simulation process that
is activated.
• Off
• Mass flow
• Volume flow
• Corrected volume
flow
• Density
• Reference density
• Temperature
Off
Value process variable One of the following
options is selected in
the
Assign process
variable
parameter:
• Mass flow
• Volume flow
• Corrected volume
flow
• Density
• Reference density
• Temperature
Enter the simulation
value for the selected
process variable.
Depends on the
process variable
selected
-
Simulation current
output 1-2
-
Switch simulation of
the current output on
and off.
• On
• Off
Off
Value current output
1-2
The
On
option is
selected in the
Simulation current
output 1-2
parameter
Enter the current value
for simulation.
3.6 to 22.5 mA
Current value
currently measured