Proline Promass E TB2 HART
Diagnostics and troubleshooting
Hauser
89
Calling up and closing the remedial measures
1.
Press
F
.
Ã
The message for the remedial measures for the selected diagnostics event opens.
2.
Press
S
+
O
simultaneously.
Ã
The message about the remedial measures closes.
• About the structure of the remedial measure message (
• For filtering the displayed event messages (
12.6.2 Filtering the event logbook
Using the
Filter options
parameter, you can define which category of event messages is displayed
in the
Events list
submenu.
Navigation path
Diagnostics
®
Event logbook
®
Filter options
Filter categories
• All
• Failure (F)
• Maintenance required (M)
• Function check (C)
• Out of specification (S)
• Information (I)
12.6.3 Overview of information events
Unlike a diagnostic event, an information event is displayed in the event logbook only and not in
the diagnose list.
Information
event
Event text
I1000
-------- (device OK)
I1079
Sensor changed
I1089
PowerOn
I1090
Configuration reset
I1091
Configuration modified
I1092
Data logging cleared
I1110
Write protection switch changed
I1111
Density adjust. error
I1137
Electronics changed
I1151
History reset
I1154
Minimum/maximum terminal voltage reset
I1155
Electronics temperature reset
I1156
Trend block memory error
I1157
Memory content events list
I1185
Device saved in display
I1186
Device with display restored
I1187
Measuring point copied via display
I1188
Display data cleared
I1189
Device backup compared
I1209
Density adjustment OK