8.3
Access to the operating menu via the local display
8.3.1
Measured value display and operating elements
Measured value display
X X X X
X X X
X
X
4
2
1
3
5
mA
20.50
F
A0013712
1
2
3
4
5
Measured value display
Device tag (
Status area
Display area for measured values (4-line)
Operating elements
Status area
The following symbols appear in the status area of the measured value display at the top right:
Status signals
A0013956
"Failure"
A device error is present. The measured value is no longer valid.
A0013957
"Maintenance required"
Maintenance is required. The measured value is still valid.
A0013959
"Function check"
The device is in service mode (e.g. during a simulation).
A0013958
"Out of specification"
The device is operated:
• Outside of its technical specifications (e.g. during startup or a cleaning)
• Outside of the configuration carried out by the user (e.g. maximum flow in parameter
20 mA
value
)
Status symbols (symbols for event level)
A0013961
"Alarm" status
The measurement is interrupted. The signal outputs take on the defined alarm condition. A
diagnostic message is generated (
A0013962
"Warning" status
The device continues to measure. A diagnostic message is generated (
Lock symbols
A0013963
Device locked
The measuring device is hardware locked (
Communication symbols
A0013965
The measuring device communicates via HART protocol.
The status symbol and status signal of the diagnostics event that currently has highest priority
for the measuring device are displayed.