Operation
Proline Promass E TB2 HART
78
Hauser
Output current 1
Measured current 1
Terminal voltage 1
Output current 2
11.4
Adapting the measuring device to the process conditions
The following are available for this purpose:
• Basic settings using the
Setup
• Advanced settings using the
Advanced setup
11.5
Perform totalizer reset
In the
Operation
submenu, 2 parameters with various options for resetting the three totalizers are
available:
• Control totalizer 1-3
• Reset all totalizers
Navigation path
"Display/operat." menu
®
Operation
"Operation" submenu
Operation
®
Control totalizer 1
Preset value 1
Control totalizer 2
Preset value 2
Control totalizer 3
Preset value 3
Reset all totalizers
Function scope of parameter "Set totalizer 1-3"
Options
Description
Reset + hold
The totaling process is stopped and the totalizer is reset to 0.
hold
The totaling process is stopped and the totalizer is set to the defined start value in the
Preset
parameter.
Reset + totalize
The totalizer is reset to 0 and the totaling process is restarted.
totalize
The totalizer is set to the defined start value in the
Preset
parameter and the totaling process
is restarted.