Proline Promass E TB2 HART
Diagnostics and troubleshooting
Hauser
87
12.5.4 Diagnostics for the configuration
Diagnostics event
Event text
Remedial measures
Event level from the
factory
F410
Data transfer
1. Check connection.
2. Retry data transfer.
Alarm
F411
Up-/Download
1. Check connection.
2. Retry data transfer.
Alarm
C411
Up-/Download active Up-/Download active, please wait.
Warning
C431
Trim
Carry out trim.
Warning
F437
Configuration
incompatible
1. Restart device.
2. Contact service.
Alarm
M438
Dataset
1. Check data set file.
2. Check device configuration.
3. Up- and download new configuration.
Warning
S441
Current output 1-2
1. Check process.
2. Check current output settings.
Warning
C453
Flow override
Deactivate flow override.
Warning
C484
Simulation failsafe
mode
Deactivate simulation.
Alarm
C485
Simulation process
variable
Deactivate simulation.
Warning
C491
Simulation current
output 1-2
Deactivate simulation.
Warning
12.5.5 Diagnostics for the process
Diagnostics event
Event text
Remedial measures
Event level from the
factory
S801
Supply voltage
Supply voltage too low!
Increase voltage.
Warning
M803
Current loop 1-2
1. Check wiring.
2. Change I/O module.
Warning
S830
Ambient temperature
range
Reduce the ambient temperature around the
sensor housing.
Warning
S831
Ambient temperature
range
Increase the ambient temperature around the
sensor housing.
Warning
S832
Ambient temperature
range
Electronics temperature is too high!
Reduce ambient temperature
Warning
S833
Ambient temperature
range
Electronics temperature is too low!
Increase ambient temperature.
Warning
S834
Process temperature
Reduce process temperature.
Warning
S835
Process temperature
Increase process temperature.
Warning
S842
Process limit value
Low flow cut off active!
Check settings for low flow cut off.
Logbook entry only
S862
Partly filled
1. Check for gas in process.
2. Adjust detection limits.
Warning
F910
Measuring tube does
not vibrate
1. Check process conditions.
2. Increase supply.
3. Check main electronic or sensor.
Alarm