Proline Promass E TB2 HART
Diagnostics and troubleshooting
Hauser
85
Diagnostics
/ ../Diagnose list
Diagnostics 3
Diagnostics 2
F273 Main electronic
A0014006-EN
Calling up and closing the remedial measures
1.
Press
F
.
Ã
The message for the remedial measures for the selected diagnostics event opens.
2.
Press
S
+
O
simultaneously.
Ã
The message about the remedial measures closes.
About the structure of the remedial measure message (
12.5
Overview of diagnostics events
Each diagnostics event is assigned a certain event level at the factory. The user can change this
assignment for certain diagnostics events.
12.5.1 Adapting event level
You can change the event level of a few diagnostics events using the
Event no. xxx
parameter.
Navigation path
"Expert" menu
®
System
®
Diagn. handling
®
Event level
®
Event no. xxx
0723-1
Event no. 044
/ ../Event level
Event no. 801
Event no. 274
Warning
A0014048-EN
You can assign the following options to the diagnostics event as the event level:
Event level
Description
Alarm
The measurement is interrupted. The signal outputs take on the defined alarm condition. A
diagnostic message is generated.
Warning
The device continues to measure. A diagnostic message is generated.
Logbook entry only
The device continues to measure. The diagnostic message is entered in the Event logbook (events
list) submenu only and not displayed in alternation with the measured value display.
Ignore
The diagnostics event is ignored, and no diagnostic message is generated or entered.
12.5.2 Diagnostics for the sensor
Diagnostics event
Event text
Remedial measures
Event level from the
factory
F022
Sensor temperature
Temperature circuit failure!
1. Change main electronic module.
2. Change sensor.
Alarm