Commissioning
Proline Promass E TB2 HART
70
Hauser
10.5.4 Configuring the totalizer
You can configure each totalizer in the three submenus
Totalizer 1-3
.
Navigation path
"Setup" menu
®
Advanced setup
®
Totalizer 1
"Setup" menu
®
Advanced setup
®
Totalizer 2
"Setup" menu
®
Advanced setup
®
Totalizer 3
Structure of the submenu
Totalizer 1-3
®
Assign process variable
Unit
Operating mode totalizer 1-3
Failure mode
Parameter overview with brief description
Parameters
Prerequisite
Description
Selection/
User entry
Factory setting
Assign process
variable
-
Select the process
variable for totalizer.
Result
The selection
determines the
picklist of the
Unit
parameter.
• Off
• Mass flow
• Volume flow
• Corrected
volume flow
Mass flow
Unit
One of the following options is
selected in the
Assign process
variable
parameter:
• Mass flow
• Volume flow
• Corrected volume flow
Select the process
variable of the
totalizer.
Unit picklist
Country-dependent:
• kg
• lb
Operation mode
totalizer 1-3
One of the following options is
selected in the
Assign process
variable
parameter:
• Mass flow
• Volume flow
• Corrected volume flow
Define how the
totalizer totalizes
the flow.
• Net flow total
• Forward flow
total
• Reverse flow total
Net flow total
Failure mode
One of the following options is
selected in the
Assign process
variable
parameter:
• Mass flow
• Volume flow
• Corrected volume flow
Define how the
totalizer behaves in
an alarm condition.
• Stop
• Actual value
• Last valid value
Stop