Technical data
Proline Promass E TB2 HART
102
Hauser
Influence of ambient
temperature
Current output (additional error, in relation to the span of 16 mA):
– Zero point (4 mA)
Average T
K
: 0.02 %/10 K, max. 0.35 % over the entire temperature range
–40 to +60 °C (–40 to +140
°
F)
– Span (20 mA)
Average T
K
: 0.05 %/10 K, max. 0.5 % over the entire temperature range
–40 to +60 °C (–40 to +140
°
F)
Influence of medium
temperature
When there is a difference between the temperature for zero point adjustment and the process
temperature, the typical measured error of the sensor is ±0.0002 % of the full scale value/
°
C
(±0.0001 % of the full scale value/
°
F).
Influence of medium pressure
The table below shows the effect on accuracy of mass flow due to a difference between calibration
pressure and process pressure.
o.r. = of reading
DN
[% o.r./bar]
[mm]
[in]
8
3/8
no influence
15
½
no influence
25
1
no influence
40
1½
no influence
50
2
–0.009
Design fundamentals
Dependent on the flow:
Flow
³
zero point stability
:
(base accuracy
:
100)
– Maximum measured error:
±
base accuracy in % o.r.
– Repeatability:
±
½
×
base accuracy in % o.r.
Flow
<
zero point stability
:
(base accuracy
:
100)
– Maximum measured error:
±
(zero point stability
:
measured value)
×
100 % o.r.
– Repeatability:
±
2
/
3
×
(zero point stability
:
measured value)
×
100 % o.r.
o.r. = of reading
Base accuracy for
[
%
o.r.]
Mass flow liquids
0.25
Volume flow liquids
0.25
Mass flow gases
0.75
18.7
Operating conditions: Installation
"Mounting requirements" chapter (
18.8
Operating conditions: Environment
Ambient temperature range
Measuring device
–40 to +60
°
C (–40 to +140
°
F)
Local display
–20 to +60
°
C (–4 to +140
°
F), the readability of the display may be impaired at
temperatures outside the temperature range.