Commissioning
Proline Promag 53 PROFIBUS DP/PA
94
Hauser
Quick Setup “Pulsating Flow”
HOME position
→
F
→
MEASURAND
→
O
→
QUICK SETUP
→
N
→
QS PULSATING FLOW (1003)
Function No.
Function name
Selection with
OS
To the next function with
F
1003
QS PULSATING FLOW
YES
After
F
is pressed by way of confirmation, the
Quick Setup menu calls up all the subsequent
functions in succession.
▼
Basic configuration
2002
DISPLAY DAMPING
3 s
Signal type for “CURRENT OUTPUT 1”
4004
MEASURING MODE
PULS. FLOW
4005
TIME CONSTANT
3 s
Signal type for “FREQ./PULSE OUTPUT 1” (for FREQUENCY operating mode)
4206
MEASURING MODE
PULS. FLOW
4208
TIME CONSTANT
0 s
Signal type for “FREQ./PULSE OUTPUT 1” (for PULSE operating mode)
4225
MEASURING MODE
PULS. FLOW
Other settings
8005
ALARM DELAY
0 s
6400
ASSIGN LOW FLOW CUTOFF
VOLUME FLOW
6402
ON-VALUE LOW FLOW CUTOFF
Recommended setting:
a0004432-en
*Full scale values
6403
OFF-VALUE LOW FLOW CUTOFF
50%
6404
PRESSURE SHOCK SUPPRESSION
0 s
▼
Back to the HOME position:
→
Press and hold down the
X
key for longer than three seconds or
→
Repeatedly press and release the
X
key
→
Exit the function matrix step by step
1000
On-value
max. full scale (per DN)*