Operation
Proline Promag 53 PROFIBUS DP/PA
70
Hauser
5
Operation
5.1
Quick operation guide
The user has a number of options for configuring and commissioning the device:
1.
Local display (option)
The local display enables you to read all important variables directly at the measuring point,
configure device-specific parameters in the field and perform commissioning.
2.
Configuration programs
The configuration of profile and device-specific parameters is primarily done via the PROFIBUS
interface. You can obtain special configuration and operating programs from various
manufacturers for these purposes.
3.
Jumpers/miniature switches for hardware settings
– PROFIBUS DP
– PROFIBUS PA
You can make the following hardware settings using a jumper or miniature switches on the I/
O board:
• Address mode configuration (select software or hardware addressing)
• Device bus address configuration (for hardware addressing)
• Hardware write protection enabling/disabling
!
Note!
A description of the configuration of the current output (active/passive) and the relay output
(NC contact/NO contact) is contained in the “Hardware settings” section.
a0001318
Fig. 43:
Methods of operating PROFIBUS PA/DP
1
Configuration/operating programs (e.g. FieldCare) for operation via PROFIBUS DP/PA
2
Jumper/miniature switches for hardware settings (write protection, device address, address mode)
3
Local display for device operation in the field (option)
4
Configuration/operating program for operating by means of the FXA 193
service interface (e.g. ToF Tool - Fieldtool Package)
1
2
3
Esc
E
-
+
Esc
E
+
-
XXX.XXX.XX
Esc
E
-
+
FXA 193
4