Proline Promag 53 PROFIBUS DP/PA
Commissioning
Hauser
115
DISPLAY_VALUE module
Any value (IEEE 754 floating point number), including status, can be cyclically transmitted via the
PROFIBUS master (Class 1) directly to the local display using the DISPLAY_VALUE module (slot 6).
Display value assignment to the main line, additional line or information line can be configured via
the local display itself or an operating program (e.g. FieldCare).
Output data
Status
The measuring device interprets the status in accordance with PROFIBUS Profile Specification
Version 3.0. The statuses OK, BAD and UNCERTAIN are indicated on the local display by a
corresponding symbol.
CONTROL_BLOCK module
By means of the CONTROL_BLOCK module (slot 7), the measuring device is able to process device-
specific control variables from the PROFIBUS master (Class 1) in cyclic data transmission (e.g.
switching on positive zero return).
Supported control variables of the CONTROL_BLOCK module
The following device-specific control variables can be activated by changing the output byte from 0
→
x:
Byte 1
Byte 2
Byte 3
Byte 4
Byte 5
Display value (IEEE 754 floating point
number)
Status
Module
Control variables
CONTROL_BLOCK
0
→
2: Positive zero return ON
0
→
3: Positive zero return OFF
0
→
8: UNIDIRECTIONAL measuring mode
0
→
9: BIDIRECTIONAL measuring mode
0
→
24: Run UNIT TO BUS function
Control variables that are executed only if the measuring device is equipped
with the optional Electrode Cleaning Circuit (ECC)
0
→
5: ECC OFF
0
→
6: ECC ON
The following control variables are only executed if the measuring device has a
relay output.
0
→
50: Relay output 1 OFF
0
→
51: Relay output 1 ON
0
→
55: Relay output 2 OFF
0
→
56: Relay output 2 ON
Control variables that are executed only if the “Batching” software plug-in is
installed in the measuring device
0
→
30: Select batching specification 1
0
→
31: Select batching specification 2
0
→
32: Select batching specification 3
0
→
33: Select batching specification 4
0
→
34: Select batching specification 5
0
→
35: Select batching specification 6
0
→
40: Stop batching
0
→
41: Start batching
0
→
42: Interrupt (pause) batching
0
→
43: Continue batching
0
→
44: Reset the fault message (affects the following batching fault messages: # 471,
472, 473, 474)
0
→
46: Reset batching counter and total batching quantity to 0