Proline Promag 53 PROFIBUS DP/PA
Troubleshooting
Hauser
141
No. # 2xx
→
Error in DAT/no communication
205
S: LOAD T-DAT
!: # 205
0x0F
BAD
Device Failure
Constant
Save to T-DAT
failed
Cause:
Data backup (downloading) to
T-DAT failed, or error when accessing (uploading)
the calibration values stored in the T-DAT.
Remedy:
1.
Check whether the T-DAT is correctly plugged
into the amplifier board.
2.
Replace the T-DAT if it is
defective.
Before replacing the DAT, check that the new,
replacement DAT is compatible with the
measuring electronics. Check the:
– Spare part set number
– Hardware revision code
3.
Replace measuring electronics boards if
necessary.
206
S: SAVE T-DAT
!: # 206
0x0F
BAD
Device Failure
Constant
Restore from T-
DAT failed
261
S: COMMUNICAT. I/O
$
: # 261
0x18
0x19
0x1A
BAD
No
Communicati
on
O.K.
Low
High
Communication
failure
Cause:
Communication error. No data reception between
amplifier and I/O board or faulty internal data
transfer.
Remedy:
Check whether the electronics boards are correctly
inserted in their holders
No. # 3xx
→
System limits exceeded
321
S: TOL. COIL CURR.
$
: # 321
0x0F
BAD
Device Failure
Constant
Coil current out of
tolerance
Cause:
The coil current of the sensor is out of tolerance.
Remedy:
1.
Remote version: Switch off the
power supply before connecting or
disconnecting the cable of the coil (terminals
41/42).
2.
Remote version:
Switch off power supply and check wiring of
terminals 41/42.
3.
Switch off the power supply and check the
connectors of the coil cable.
4.
Replace measuring electronics boards if
necessary.
No.
Device status message
(local display)
PROFIBUS measured value status
Extended
diagnostic
message in the
PROFIBUS
master
Cause/remedy
Quality code (HEX)
Measured value status
Quality status
Quality substatus
Limits