Operation
Proline Promag 53 PROFIBUS DP/PA
72
Hauser
5.2.2
Display (operating mode)
The display area consists of three lines in all; this is where measured values are displayed, and/or
status variables (direction of flow, bar graph, etc.). You can change the assignment of display lines
to different variables to suit your needs and preferences (
→
see the “Description of Device
Functions” manual).
Multiplex mode:
A maximum of two different display variables can be assigned to each line. Variables multiplexed in
this way alternate every 10 seconds on the display.
Error messages:
Display and presentation of system/process errors
a0004755
Fig. 45:
Typical display for normal operating mode (HOME position)
1
Main display line: shows primary measured values, e.g. mass flow in [kg/h]
2
Additional line: shows measured variables and status variables, e.g. totalizer No. 3 in [t]
3
Information line: shows additional information on the measured variables and status variables,
e.g. bar graph display of the full scale value achieved by the mass flow
4
“Info icons” field: icons representing additional information on the measured values are shown in this field.
5
“Measured values” field: the current measured values appear in this field
6
“Unit of measure” field: the units of measure and time defined for the current measured values appear in this field
5.2.3
Additional display functions
Depending on the order option (F-CHIP*), the local display has different display functions:
Device without batching software:
From HOME position, use the
OS
keys to open an “Info Menu” containing the following
information:
• Totalizer (including overflow)
• Actual values or states of the configured inputs/outputs
• Device TAG number (user-definable)
OS
→
Scan of individual values within the Info Menu
X
(Esc key)
→
Back to HOME position
Device with batching software:
On measuring devices with installed batching software (F-Chip*) and a suitably configured display
line, you can carry out filling processes directly using the local display. You will find a detailed
description on
1
4
5
6
2
3
+24.502
+1863.97
x
x
y
y
–50
+50
%
v
v
TOT
OK
3
*F-CHIP