Proline Promag 53 PROFIBUS DP/PA
Wiring
Hauser
57
4.2.2
Cable specifications
Coil cable
• 2 x 0.75 mm
2
PVC cable with common, braided copper shield (
∅
∼
7 mm)
• Conductor resistance:
≤
37
Ω
/km
• Capacitance: core/core, shield grounded:
≤
120 pF/m
• Permanent operating temperature: –20 to +80 °C
• Cable cross-section: max. 2.5 mm
2
Signal cable
• 3 x 0.38 mm
2
PVC cable with common, braided copper shield (
∅
∼
7 mm) and individually
shielded cores
• With Empty Pipe Detection (EPD): 4 x 0.38 mm
2
PVC cable with common, braided copper shield
(
∅
∼
7 mm) and individually shielded cores
• Conductor resistance:
≤
50
Ω
/km
• Capacitance core/shield:
≤
420 pF/m
• Permanent operating temperature: –20 to +80 °C
• Cable cross-section: max. 2.5 mm
2
a0003194
Fig. 33:
Cable cross-section
a
Signal cable
b
Coil current cable
1
Core
2
Core insulation
3
Core shield
4
Core jacket
5
Core reinforcement
6
Cable shield
7
Outer jacket
As an option, Hauser can also deliver reinforced connecting cables with an additional,
reinforcing metal braid. We recommend such cables for the following cases:
• Directly buried cable
• Cables endangered by rodents
• Device operation which should comply with the IP 68 standard of protection
Operation in zones of severe electrical interference
The measuring device complies with the general safety requirements in accordance with
EN 61010, the EMC requirements of EN 61326/A1, and NAMUR recommendation NE 21.
"
Caution!
Grounding is by means of the ground terminals provided for the purpose inside the connection
housing. Keep the stripped and twisted lengths of cable shield to the terminals as short as possible.